Cooling, Ac power source, Power and ground connections – Teledyne LeCroy SAS_SATA InFusion - Users Manual User Manual
Page 12

Version 2.0
SAS InFusion User Manual
xii
LeCroy Corporation
Cooling
The SAS/SATA InFusion relies on forced air cooling with internal fans and ventilation 
openings. Care must be taken to avoid restricting the airflow around the apertures (fan 
holes). To ensure adequate ventilation, it is required to leave a 10 cm (4 inch) minimum 
gap around the sides of the instrument.
AC Power Source
The SAS/SATA InFusion operates from a single-phase, 100-240 Vrms (+/-10%) AC 
power source at 50/60 Hz (+/-5%). 
No manual voltage selection is required because the instrument automatically adapts to 
line voltage.
Depending on the accessories installed (PC port plug-ins, etc.), the instrument can draw 
up to 55 W (55 VA).
The power supply of the SAS/SATA InFusion is protected against short circuit and 
overload by a 5x20 mm fuse (T1.0 A/250 V). See "Fuse Replacement" in this section for 
replacement instructions.
Power and Ground Connections
The SAS/SATA InFusion is provided with a grounded cord set containing a molded 
three-terminal polarized plug and a standard IEC320 (Type C13) connector for making 
line voltage and safety ground connection. The AC inlet ground terminal is connected 
directly to the frame of the instrument. For adequate protection against electrical shock 
hazard, the power cord plug must be inserted into a mating AC outlet containing a safety 
ground contact.
CAUTION
Do not block ventilation holes located on both sides of
SAS/SATA InFusion
.
NOTE:
The SASInFusion/SATAInFusion automatically adapts itself to the AC 
line input within the following ranges:
Voltage Range: 90 - 264 Vrms.
Frequency Range: 47-63 Hz.
WARNING
Electrical Shock Hazard!
Any interruption of the protective conductor inside or outside 
of the 
SAS/SATA InFusion
, or disconnection of the safety
ground terminal creates a hazardous situation.
Intentional interruption is prohibited.