Rio setup, Up down select back – Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 98
93
The user will be prompted:
“Short the Sim. Cable Leads (clip together)”
Clip the alligator clips together and then Select the
OK soft key. The RIO4000 will perform a self
calibration and then the user will be prompted:
“Connect Sim. Cable Leads to the device”
Below is an example of 2, 3, and 4-Wire RTD Configurations for an RTD Sensor to a given
manufacturer’s Temperature Transmitter. There can be different configurations, however these
examples illustrate the corresponding RIO4000 connection required to perform RTD Simulation for 2, 3
and 4 Wire RTD’s.
RIO Setup
Settings & Functions
Choose New Probe
Display: Degrees or
Ω
Simulate Mode
Measure Mode
Selected Probe:
S->Pt 100 (385)
Up Down Select Back
1
2
3
4
Sensor Connections
JUMPER
Example 4-Wire RTD
Configuration.
1
2
3
4
RIO4000
Example 3-Wire RTD
Configuration.
Example 2-Wire RTD
Configuration.
1
2
3
4
1
2
3
4
1
2
3
4
RIO4000
JUMPER
1
2
3
4
RIO4000
JUMPER
RIO4000 Connections