beautypg.com

Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 61

background image

56

FUNCTION and then the PgDn soft key. To retreat to the prior page, press FUNCTION and then PgUp.
Once engaged, the PgUp and PgDn soft keys remain active until the Up or Down soft keys or FUNCTION
is pressed again.

List / Edit Configs:
Lists all stored HART configurations by Tag Number. Once a
desired tag is located with the cursor, pressing the Select soft key
displays header information to help the user confirm the identity
and origin of the configuration. Pressing Select again opens the
configuration for review and editing as needed. Any configuration
in the list can be sent to another HART device of the same
manufacture and model number. This feature enables fast
commissioning of replacement devices, cloning of existing
systems, or re-configuring for changes in process conditions or
batch runs. New entries to this list are loaded to the bottom of the
list. If multiple entries for one device are shown, the most recently
saved configuration is always listed last.

Entries in this list may have been saved directly from a connected HART device, created in the MFT using
the Create Configs function, copied from another configuration and renamed, or created by editing an
existing file. MFT will also list configurations downloaded from the separately available Meriam Device
Management System software (see www.meriam.com for more information).

Create Configs:
The Create Configs function allows the user to configure a HART device file in Offline Mode for sending to
the intended device later when connected in the Online Mode. Use the soft key controls to move through
the list and select the device model number required. The menus prompt the user to make the necessary
configuration selections, tag the file for later retrieval and use, and edit configuration lines as required. This
function can be locked out. See the Lockout Section of this manual for more information.

Delete Individual Configs / Clearing Configuration Memory:
Individual stored Configurations may be deleted with MFT Del soft
key functions (see example at right). A delete confirmation screen is
provided to prevent unintended deletions. The Del key only hides the
record from the MFT screen; it does not clear memory space.

To clear all configurations from memory (including active and hidden
configs), enter the HART Offline Menu Screen and select the
Delete All Configs option. The MFT will ask “Delete ALL Configs…
ARE YOU SURE???” Select Yes to clear all stored configurations
from memory.



HART ONLINE MODE
Once a HART device is located, the MFT enters the Hart Device Online Screen as shown below. The word
“Hart” appears in the top left of LCD. The symbol will “flash” when HART communications have been
established.

Note: The screen at right illustrates a Rosemount 1151 rev.
5 HART display. Screen configuration may vary depending
on the HART device connected. For more information,
please refer to the owner’s manual for the HART device
under test.






Hart
1151: Pump 8

Device Config

Pres: 0.000 InH20
AO:

4.000 mA

LRV:

0.000 InH20

URV: 104.8 InH20
% of Range: 0.000%

AO Fixed

Up Down Select Back

Offline List / Edit

Configs Screen

Offline

List/Edit Configs

PT-105B
DPT-201C
TT-312 A

TT-312 A
FE-201
FE-201
PT-6174

Up Down Select Back

Delete Individual Configuration

Offline

List/Edit Configs
PT-105A

Device ID 6121486
Single Var
Endress + Hauser
Cerabar S
Complete Config
Uses DOF 11070702
81 Variables
02/21/03 4:50 PM

Copy Del Select Back

This manual is related to the following products: