S3: 0.00 psig, Iv: 0.000 ma, S1: 72.8 deg f – Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 87: T/c open
82
TIO0110 Module Quick Instructions
For Thermocouple and milli-Volt Measurement and Simulation
Safety Notice: Intrinsically Safe TIO0110 modules require the MFT Intrinsically Safe base units. The TIO module
must be installed or removed in a safe area. T/C connectors must be connected to the TIO0110 module in a non-
hazardous area. For more information on Intrinsically Safe operation, see the MFT User’s Manual section on this
topic and see the Intrinsic Safety Control Document in the appendix.
The TIO0110 Module will measure any of twelve (12) thermocouple types, measure milli-Volts, or simulate the milli-
Volt output the twelve (12) t/c types. Connections for measurement applications are made using mini-t/c connectors
with appropriate t/c wire. Connections for simulation applications are made using mini-t/c connectors with wire cable
and alligator clip ends (see Accessories section for the Meriam wire-kit part numbers).
Measure
• T/C types E, J, K, N, T
• T/C types B, K, R
• T/C types C, D, S, M, P
• ± 110.000 mV dc
1.0 Measurement Mode
1.1 Install the TIO0110 module in any MFT bay location and turn on the MFT. If no t/c is connected to the
MFT, the Sx line will indicate “T/C open.” If a t/c is connected a temperature will be shown on the display.
Be sure the t/c type is correct - if not go to the TC Setup display to change the type.
1.2 To access the setup menu for the TIO0110 module, press the
More soft key from the Measure Mode of
the MFT and then the
Sensr soft key to view the S1 sensor bay module information. If the TIO module
information is shown, press the
Setup soft key to arrive at the TC Setup display shown below. Otherwise,
press the
Next soft key until the TIO module is displayed.
Simulate
• ± 110.000 mV dc output
for the same t/c types
MFT Measure Mode Display showing open T/C
Measure
S1:
T/C open
S2: 0.000 inW20C
S3: 0.00 PSIG
IV: 0.000 mA
Zero Mn/Mx Damp More
MFT Measure Mode Display showing temperature
Measure
S1: 72.8 deg F
S2: 0.000 inW20C
S3: 0.00 PSIG
IV: 0.000 mA
Zero Mn/Mx Damp More
Sensor
Sensor 1
MERIAM
Model:
TIO0110
SN:
xxxxxxxxx
Cal Date:
YYYY-MM-DD
Next Setup
TC Setup
Current Settings:
TC Type
J
Standard
ITS 90
Resolution
0.1 deg
Simulate Mode
User Re-Calibrate
Up Down Select Back