beautypg.com

S3: 0.00 psig, Iv: 0.000 ma, S1: 72.8 deg f – Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 87: T/c open

background image

82

TIO0110 Module Quick Instructions

For Thermocouple and milli-Volt Measurement and Simulation
Safety Notice:
Intrinsically Safe TIO0110 modules require the MFT Intrinsically Safe base units. The TIO module
must be installed or removed in a safe area. T/C connectors must be connected to the TIO0110 module in a non-
hazardous area. For more information on Intrinsically Safe operation, see the MFT User’s Manual section on this
topic and see the Intrinsic Safety Control Document in the appendix.

The TIO0110 Module will measure any of twelve (12) thermocouple types, measure milli-Volts, or simulate the milli-
Volt output the twelve (12) t/c types. Connections for measurement applications are made using mini-t/c connectors
with appropriate t/c wire. Connections for simulation applications are made using mini-t/c connectors with wire cable
and alligator clip ends (see Accessories section for the Meriam wire-kit part numbers).

Measure

• T/C types E, J, K, N, T

• T/C types B, K, R

• T/C types C, D, S, M, P
• ± 110.000 mV dc


1.0 Measurement Mode

1.1 Install the TIO0110 module in any MFT bay location and turn on the MFT. If no t/c is connected to the

MFT, the Sx line will indicate “T/C open.” If a t/c is connected a temperature will be shown on the display.
Be sure the t/c type is correct - if not go to the TC Setup display to change the type.

















1.2 To access the setup menu for the TIO0110 module, press the

More soft key from the Measure Mode of

the MFT and then the

Sensr soft key to view the S1 sensor bay module information. If the TIO module

information is shown, press the

Setup soft key to arrive at the TC Setup display shown below. Otherwise,

press the

Next soft key until the TIO module is displayed.












Simulate
• ± 110.000 mV dc output

for the same t/c types

MFT Measure Mode Display showing open T/C

Measure

S1:

T/C open

S2: 0.000 inW20C

S3: 0.00 PSIG

IV: 0.000 mA



Zero Mn/Mx Damp More

MFT Measure Mode Display showing temperature

Measure

S1: 72.8 deg F

S2: 0.000 inW20C

S3: 0.00 PSIG

IV: 0.000 mA



Zero Mn/Mx Damp More

Sensor
Sensor 1
MERIAM
Model:

TIO0110

SN:

xxxxxxxxx

Cal Date:

YYYY-MM-DD






Next Setup

TC Setup
Current Settings:
TC Type

J

Standard

ITS 90

Resolution

0.1 deg

Simulate Mode
User Re-Calibrate




Up Down Select Back