Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 53
48
Select Begin Calibration once satisfied. The following screens appears:
Page 1
Page 2 (
Use the MFT user interface to enter the Tag, S/N, and other information needed. Be sure the
Meas/Sim line is directed to the RIO4000 module location (S1, S2 or S3). Use Temperature Setup
on the bottom of the second page to setup or change the RTD probe information (units, number of
wires, RTD material, alpha coefficient, etc…). When satisfied, move the cursor arrow to the Begin
Calibration menu and press the Select soft key to continue. The following screen will appear.
Select the number of set points for the switch device of interest and press the Select soft key to
continue. The following screens will be displayed. Use this MFT user interface to enter the rest of the
switch test detail. Be sure to consult the information on the
specify a LRV and URV for the test. This is especially useful for switch tests using a simulation
module since these values will provide start and stop values for the simulation module.
Cal
Non-Hart Cal
Begin
Calibration
Tag:
DEFAULT
S/N:
00000
Meas/Sim: S2
LRV : 0.000
URV : 10.000
Units : C
Xfer Fn : Linear
Up Down Select Back
Cal
Non-Hart Cal
Begin
Calibration
Measure: IV
Output LRV: 19.900
Output URV: 20.000
Units:
V
Temperature Setup
Up Down Select Back
Cal
Number of Set Points:
1
2
3
4
5
6
Choose the number
of Set / Reset PAIRS
Up Down Select Back
Switch Test
Setup: As Found
Start 1 of 1
Set Pt : 7.000
Reset Pt: 6.5.000
Direction : Inc
Test Reset Pt : Yes
Ramp Minutes: 5
Start Delay : 0
Contact Form : NO
VMA (wetting) : On
Up Down Select Back
Page 1
Page 2 (
Switch Test
Setup: As Found
Start 1 of 1
LRV : 0.000
URV : 10.000
Up Down Select Back