Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 85
80
Sensor Modules List
For a precise list of Intrinsic Safety certified modules, please refer to the Intrinsic Safety
Control Drawing in the Appendix of this manual.
FDN0010
1 - 10" H20 dp
± 0.1% FS from 23° F to 122° F (± 0.009 H20 dp)
FDN0020
0 - 20" H20 dp
± 0.1% FS from 23° F to 122° F (± 0.020 H20 dp)
DDN0100
0 - 100" H20 dp
± 0.025 " H20 dp
DDN0200
0 - 200" H20 dp
± 0.05 " H20 dp
DDN0400
0 - 400" H20 dp
± 0.10 " H20 dp
DDN2000
0 - 2000" H20 dp
± 0.50 " H20 dp
DGI0020
0 - 20 PSIG
± 0.005 " PSIG
DGI0200
0 - 200 PSIG
± 0.05 " PSIG
DGI0500
0 - 500 PSIG
± 0.125 " PSIG
DGI1000
0 - 1000 PSIG
± 0.25 " PSIG
DGI1500
0 - 1500 PSIG
± 0.375 " PSIG
DGI2000
0 - 2000 PSIG
± 0.500 " PSIG
DAI0900
0 - 900mm Hg Abs
± 0.225 mm Hg Abs
DAI2000
0 - 2000mm Hg Abs
± 0.50 mm Hg Abs
DAI5200
0 - 5200mm Hg Abs
± 13.00 mm Hg Abs
GGI3000
0 - 3000 PSIG
± 0.05% FS from 23° F to 122° F (± 1.50 PSIG)
TIO0110
Starter Kit for TIO0110 module t/c meas. & sim. - Universal
T/C wire Kit 1 for E, J, K, N, T; mini-t/c conn., 1 m cable
t/c wire Kit 1 for B, K, R/S; mini-t/c conn., 1 m cable
mV meas. / sim. cable assbly; mini-t/c conn., 1 m cable
VMA0055
Starter Kit for VMA0055 module V /
I
meas. & sim. - Universal
RIO4000
Starter Kit for RIO4000 module RTD meas. & sim. - Universal
Measure cable; module connector by four (4) alligator clips
RTD mating connector (for use with customer's RTD)
Simulate Cable; module connector by two (2) aligator clips
V/I: Measure ±55 Vdc, Source 0-28Vdc; mA: Measure ±100 mA DC,
Source 0-100 mA dc (Recommend VMA Test Lead Kit (P/N
VMA test lead kit, banana plugs on 9" breakouts both ends, assorted
connectors
RTD measure and simulate module supports 22 different RTDs
V/mA Meas./Sim.
RTD Meas./Sim.
PRESSURE SENSOR MODULE MODEL NUMBER, RANGE & ACCURACY
Modules are Intrinsically Safe when used in accordance with Control Doc.
"D" Series Accuracy is: ± 0.025% FS from 23° F to 122° F
T/C measure and simulate module supports types
B,C,E,J,K,N,R,S,T,M,P
Measure and Simulation Module Model Number (
Modules are Intrinsically Safe when used in accordance with
Control Doc).
T/C Meas./Sim.