beautypg.com

Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 103

background image

98














2.2.2 If an RTD type is connected (depending on the Rs value), the MFT displays the measured temperature.

Units can be changed by pressing the MFT

Function key (and letting go) and then the appropriate Units

Sensor X key. The various temperature units will display after each button push. Press the Function
key again to return the MFT to standard Measure Mode.

2.2.3 When the MFT is in Measure Mode and the User Measure Cable is connected to the RIO4000 (Rs = 0 )

the MFT will default to the last user setting for RTD Type and display the measured temperature and/or
Ohm reading.
Note:

The last user setting for RTD Type is briefly displayed at the bottom of the LCD.

2.2.4 To Select a New RTD Type, Select the

More soft key, then the Sensr soft key and choose the correct

“Sensor X” location that corresponds
with the module bay that the RIO
module and select the

Setup soft

key. The following will be displayed:









2.2.5 Move the right arrow to the “Select New RTD Type” option and press the Select soft key. The following

will be displayed.
















Measure

S1: No probe

S2: 0.000 inW20C

S3: 0.00 PSIG

IV: 0.000 mA


Zero Mn/Mx Damp More

MFT Measure Mode Display showing open RTD

MFT Measure Mode Display showing temperature

Measure

S1: 72.8 deg F

S2: 0.000 inW20C

S3: 0.00 PSIG

IV: 0.000 mA


Zero Mn/Mx Damp More

RIO User Select

Select New RTD Type
Pt 100 (385)
Pt 1000 (385)
Pt 100 (392)
Pt 1000 (391)
Pt 25.5 (392)
Pt 200 (385)
Pt 470 (392)
Pt 500 (392)
Pt 500 (391)
Pt 500 (385)

Up Down Select Back

RIO User Select

Select New RTD Type
Fe 1000 (527)
Fe 2000 (527)
User Select
1000

Cal Res

100

Cal Res

10

Cal Res








Up Down Select Back

RIO User Select

Select New RTD Type
Pt 1000 (375)
Pt 100 (393)
Pt 100 (389)
Pt 98 (392)
Pt 200 (392)
Cu 10 (427)
Cu 100 (427)
Ni 100 (618)
Ni 120 (672)
Fe 604 (518)

Up Down Select Back

RIO User Select screen 1

RIO User Select screen 3

RIO User Select screen 2

RIO Setup

Settings & Functions

Select New RTD Type
Display: Degrees or

Measure Mode
Simulate Mode


Selected RTD Type:
M->Pt 100 (385)


Up Down Select Back

This manual is related to the following products: