Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 30
25
Select the Function menu option under the Meas/Sim line to select the needed MFT operation for the
device under test’s input. The following screen is shown:
P Meas: is selected for P transmitter, P/I converter, P/P converter, or P switch calibrations.
T/C Meas: is selected for t/c comparison testing or bulb type T switch calibrations
RTD Meas: is selected for RTD comparison testing or bulb type T switch calibrations.
VMA Meas: is generally not used
T/C Sim: is selected for T/C temperature transmitter calibration
RTD Sim: is selected for RTD T transmitter calibrations or RTD switches.
VMA Sim: is selected for I/P converter calibrations and some special applications
Press the Back soft key if you choose not to make any changes. The MFT will then return to
the Calibrate key main menu screen.
Select the Location menu option under the Meas/Sim line to select the sensor needed for the
Meas/Sim type displayed (previously selected). The following screen is shown.
Select the Function menu option under the Measure line to select the needed MFT operation for the
device under test’s output. The following screen is shown:
Cal
Meas/Sim: P Meas
S1: RIO
S2: DDN
S3: TIO
Up Down Select Back
Cal
Meas/Sim: P Meas
P
Meas
TC
Meas
RTD Meas
VMA Meas
TC Sim
RTD Sim
VMA Sim
Up Down Select Back
Cal
Meas: IV Meas
P
Meas
TC
Meas
RTD Meas
VMA Meas
IV Meas
Up Down Select Back