beautypg.com

Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 101

background image

96

Table 1: Supported RTD Materials / Rs Sensor Resistor No.

Material R0

ALPHA

Rs

(Probe ID)

Temp Range
(deg C)

Recommended
Calib Resistor

Platinum 100

0.00385

100

-200 to +850

° C

100 Ω

Platinum 1000

0.00385

1000

-200 to +630

° C

1000 Ω

Platinum 100

0.00392

46.4

-200 to +630

° C

100 Ω

Platinum 100

0.00391

56.2

-200 to +630

° C

100 Ω

Platinum 25.5

0.00392

17.8

-200 to +630

° C

100 Ω

Platinum 200

0.00385

178

-200 to +630

° C

100 Ω

Platinum 470

0.00392

261

-200 to +630

° C

1000 Ω

Platinum 500

0.00392

316

-200 to +630

° C

1000 Ω

Platinum 500

0.00391

383

-200 to +630

° C

1000 Ω

Platinum 500

0.00385

464

-200 to +630

° C

1000 Ω

Platinum 1000

0.00375

1210

-50 to +500

° C

1000 Ω

Platinum 100

0.00393

68.1

-200 to +962

° C

100 Ω

Platinum 100

0.00389

82.5

-200 to +630

° C

100 Ω

Platinum 98.129

0.00392

21.5

-200 to +600

° C

100 Ω

Platinum 200

0.00392

215

-200 to +630

° C

100 Ω

Copper 9.035

0.00427

12.1

-100 to +260

° C

10 Ω

Copper 100

0.00427

121

-100 to +260

° C

100 Ω

Nickel 100

0.00618

31.6

-60 to +160

° C

100 Ω

Nickel 120

0.00672

147

-80 to +260

° C

100 Ω

Iron 604

0.00518

562

-100 to +200

° C

1000 Ω

Iron 908.4

0.00527

681

-100 to +200

° C

1000 Ω

Iron 1816.81

0.00527

1470

-100 to +200

° C

1000 Ω

10 Ω resistor

na

na

14.7

Precision cal resistor

na

100 Ω resistor

na

na

26.1

Precision cal resistor

na

1000 Ω resistor

na

na

825

Precision cal resistor

na

Based on the Rs value connected between pins 1 and 2, the corresponding RTD Type look up table is
utilized.

Plug connectors are available for customers that want to make up their own RTD Type standards. If no
Rs resistor is installed between pins 1 and 2, a copper wire (Rs = 0) must be wired between pins 1 and 2.
When Rs = 0 is detected, the RTD Type setting will be defaulted to the last RTD Types setting used.
This setting will display briefly at the bottom of the MFT display when the cable is attached to the RIO
module.

Option 1: RTD wiring styles for User Wired RTD Type with Manufacturer’s Plug

IMPORTANT: Pins 3 – 6 on the custom connector must all be connected for readings to take place!

2-wire

RTDs

Connect pins 3 and 4 to one wire/terminal of the

RTD. Connect pins 5 and 6 to the other wire/terminal.


3-wire

RTDs

3-wire RTDs have two wires/terminals that are the same

color. Connect pin 3 to one of these wires/terminals and

connect pin 4 to the other wire/terminal of the same color.

Connect both pins 5 and 6 to the remaining, different

color

wire/terminal.


4-wire

RTDs

4-wire RTDs have two wires/terminals of one color and two

of another color. Connect pins 3 and 4 to each of two

wires/terminals of the same color. Connect pins 5 and 6

to each of two wires/terminals of the other color.