Table 1: Supported RTD Materials / Rs Sensor Resistor No.
Material R0
ALPHA
Rs
(Probe ID)
Temp Range
(deg C)
Recommended
Calib Resistor
Platinum 100
0.00385
100
-200 to +850
° C
100 Ω
Platinum 1000
0.00385
1000
-200 to +630
° C
1000 Ω
Platinum 100
0.00392
46.4
-200 to +630
° C
100 Ω
Platinum 100
0.00391
56.2
-200 to +630
° C
100 Ω
Platinum 25.5
0.00392
17.8
-200 to +630
° C
100 Ω
Platinum 200
0.00385
178
-200 to +630
° C
100 Ω
Platinum 470
0.00392
261
-200 to +630
° C
1000 Ω
Platinum 500
0.00392
316
-200 to +630
° C
1000 Ω
Platinum 500
0.00391
383
-200 to +630
° C
1000 Ω
Platinum 500
0.00385
464
-200 to +630
° C
1000 Ω
Platinum 1000
0.00375
1210
-50 to +500
° C
1000 Ω
Platinum 100
0.00393
68.1
-200 to +962
° C
100 Ω
Platinum 100
0.00389
82.5
-200 to +630
° C
100 Ω
Platinum 98.129
0.00392
21.5
-200 to +600
° C
100 Ω
Platinum 200
0.00392
215
-200 to +630
° C
100 Ω
Copper 9.035
0.00427
12.1
-100 to +260
° C
10 Ω
Copper 100
0.00427
121
-100 to +260
° C
100 Ω
Nickel 100
0.00618
31.6
-60 to +160
° C
100 Ω
Nickel 120
0.00672
147
-80 to +260
° C
100 Ω
Iron 604
0.00518
562
-100 to +200
° C
1000 Ω
Iron 908.4
0.00527
681
-100 to +200
° C
1000 Ω
Iron 1816.81
0.00527
1470
-100 to +200
° C
1000 Ω
10 Ω resistor
na
na
14.7
Precision cal resistor
na
100 Ω resistor
na
na
26.1
Precision cal resistor
na
1000 Ω resistor
na
na
825
Precision cal resistor
na
Based on the Rs value connected between pins 1 and 2, the corresponding RTD Type look up table is
utilized.
Plug connectors are available for customers that want to make up their own RTD Type standards. If no
Rs resistor is installed between pins 1 and 2, a copper wire (Rs = 0) must be wired between pins 1 and 2.
When Rs = 0 is detected, the RTD Type setting will be defaulted to the last RTD Types setting used.
This setting will display briefly at the bottom of the MFT display when the cable is attached to the RIO
module.
Option 1: RTD wiring styles for User Wired RTD Type with Manufacturer’s Plug
IMPORTANT: Pins 3 – 6 on the custom connector must all be connected for readings to take place!
2-wire
RTDs
Connect pins 3 and 4 to one wire/terminal of the
RTD. Connect pins 5 and 6 to the other wire/terminal.
3-wire
RTDs
3-wire RTDs have two wires/terminals that are the same
color. Connect pin 3 to one of these wires/terminals and
connect pin 4 to the other wire/terminal of the same color.
Connect both pins 5 and 6 to the remaining, different
color
wire/terminal.
4-wire
RTDs
4-wire RTDs have two wires/terminals of one color and two
of another color. Connect pins 3 and 4 to each of two
wires/terminals of the same color. Connect pins 5 and 6
to each of two wires/terminals of the other color.