beautypg.com

Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 109

background image

104


RIO4000 Accuracy Table

Ambient temperature limits: -10 ºC to +50 ºC

Accuracy statements are for ambient temperatures of 18º C to 28º C.
Apply the Temperature Coefficient for ambient temperatures below 18º C and above 28º C

RTD Type

(α)

T Range (˚C) Measure (2, 3, 4 Wire)

Simulate (2 Wire**)

Pt 100

385

-200 to 0

0.3

0.3

0 to 400

0.4

0.4

400 to 800

0.5

0.6

Pt 100

391

-200 to 0

0.3

0.4

0 to 630

0.4

0.6

392

-200 to 0

0.4

0.4

0 to 630

0.5

0.6

Pt 100

393

-200 to 0

0.5

0.4

0 to 400

0.4

0.5

400 to 962

0.6

0.6

Pt 200

385, 392

-200 to 0

0.2

0.5

0 to 400

0.6

0.4

400 to 630

0.8

0.6

Pt 500

385, 392

-200 to 0

0.2

0.4

0 to 400

0.3

0.5

400 to 630

0.4

0.6

Pt 500

391

-200 to 0

0.2

0.5

0 to 400

0.4

0.4

400 to 630

0.5

0.6

Pt 1000

385

-200 to 0

0.2

0.5

0 to 400

0.3

0.4

400 to 630

0.4

0.6

Pt 1000

375

-50 to 0

0.3

0.4

0 to 500

0.3

0.5

Cu 10

4274

-100 to 0

2

2

0 to 260

2

2

Cu 100

4274

-100 to 0

1

1

0 to 260

1

1

Ni 100

618

-60 to 160

0.2

0.4

Ni 120

672

-80 to 260

0.2

0.4

Fe 604

518

-100 to 200

0.3

0.5

Fe 908

527

-100 to 200

0.3

0.5

Fe 1816

527

-100 to 200

0.2

0.5

*Temperature Coefficient = 0.02 ˚C/˚C
** For 2-Wire Simulation, Manufacturer's 2 Wire Connector/Cable is Required.

Operating Temperature 18

This manual is related to the following products: