Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 82
77
ELECTRICAL MEASUREMENTS
Current: ±50.00 mA
±0.025% FS ± 0.001% of F.S. per ° F (ref temp is 70° F)
Voltage: ±50 VDC
±0.025% FS ± 0.001% of F.S. per ° F (ref temp is 70° F)
MODULES: 2.3”L x 1.3”W x 2.6”D, 8.3 ozs. Positive-lock with spring-loaded sensor lock. Depress lock to release.
MIN/MAX - minimum and maximum measured values are captured. Capture speed equal to current damping rate
setting.
HOLD - toggle function freezes display.
ZERO – simultaneous re-zeroing of all installed modules
CONTRAST – direct adjustment of display contrast
FIELD RECERTIFICATION – sensor modules and integral mA / Volts can be recalibrated for zero, span and linearity
through the base unit using appropriate standards at 1, 3, 5, or 9 points (exception: VMA0055 module must be
returned to factory for recalibration).
SECURITY LOCKOUT – MFT 4000 Settings can be locked out entirely or by specific feature.
CE compliance