83
1.3 At the TC Setup display the user can select the desired t/c from the TC Type list, select the mV to
temperature conversion standard, set the temperature display resolution, select Simulate Mode or enter
the User Re-Calibrate Mode. Arrow Up or Down to the desired menu option and press the Select soft
key.
1.3.1
TC Type – select from T/C types E, J, K, N, T, B, R, C, D, S, M, or P. Use the Up and Down
soft keys and Save key to set type.
1.3.2
Standard – choose from NIST 175 (default), ITS 90 or ITS 68
standards. Use the Up and Down soft keys and Save key to set.
1.3.3
Resolution – use to select temperature display resolution of 1
° or 0.1°. Use the Up and
Down soft keys and Save key to set.
1.3.4
Simulate Mode – select simulate mode to simulate a t/c mV output to a receiving device.
See details below.
1.3.5
User Re-Calibration – select this menu option to calibrate the TIO0110 module with
appropriate local standards. See details below.
2.0 Simulate Mode
Select
Simulate Mode from the TC Setup screen to access the TC simulation display. From this display the user
can select simulation units, start a simple simulation or setup and execute an automated test procedure.
2.1 The
Start menu begins a TC or mV simulation. When Start is selected with the Select soft key, a numeric
entry screen appears to facilitate the entry of the desired simulation temperature value (or mV value).
Once the value has been entered, the MFT will use the previously selected conversion standard to
determine the corresponding mV value and output that value through the TIO mini-t/c connector.
2.2 The
Units menu line is used to select the temperature units for the simulation.
°C, °F, °K, or °R are the
available temperature units. The user can also select mV units for the simulation if desired.
2.3 Step Value is a user entered value that can be used to increment or decrement the simulation value after
the simulation is started.
2.4
Auto-Step x menu options allow pre-programmed simulations to be set up and executed automatically by
the MFT / TIO. Auto-Steps are automatically stored for future use.
.
2.4.1 Select
Start menu option to begin the indicated Auto-Step number.
2.4.2 Select
Units to change the Auto-Step x simulation units.
TC Simulation
Simulation Mode
Start
Units:
C
Step Value 10 C
Auto Step 1
Auto Step 2
Auto Step 3
Auto Ramp 1
Auto Ramp 2
Auto Ramp 3
Up Down Select Back
TC Setup
Auto-Step 1
Start
Units: C
Low pt: 100.0
Upr pt: 400.0
Steps: 3
Dwell Time: Manual
Up Down Select Back