Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual
Page 95
90
Start – Begins the desired simulation mode with the stepping
parameters selected.
Mode – Displays the current selected simulation mode. Select
Mode to change.
Low pt – Displays the starting source or simulation value.
Select to enter a value.
Upr pt – Displays the ending source or simulation value. Select
to enter a value.
Steps – Displays the number of steps to take (including the
“Upr pt”). Select to change the number of steps.
Dwell Time – Displays the time delay in seconds at each step.
“Manual” can be selected to enable user to
manually advance to next programmed step.
5.0 Resolution, Range, Accuracy for VMA0055 Module
mA dc
mA Measure Mode
mA Source Mode
Resolution
.001 mA
.001 mA
Range
±100.000 mA (55 Vdc compliance)
0 - 22.000 mA
Accuracy
± (0.01% of reading + 0.015% FS)
± (0.01% of reading + 0.015% FS)
Open Circuit Voltage
24 VDC
Output Drive
15 VDC minimum @ 24 mA, Resistive load
Volts dc
V dc Measure Mode
V dc Source (Regulated) Mode
Resolution
1/100,000 counts: .001 mV; .001 V
1 / 1,000,000
Ranges
500 mV; 1, 2, 4, 8, 15, 30, 55 V
± 1000 mV, 0 - 24.000 VDC
Accuracy
± (0.01% of reading + 0.05% FS)
± (0.01% of reading + 0.05% FS)
Open Circuit Voltage
24 VDC
Output Drive
15 VDC minimum @ 24 mA, Resistive load
2-wireTransmitter Simulate
mA Simulation
Resolution
.001 mA
Range
0 - 24.000 mA
Accuracy
± (0.01% of reading + 0.015% FS)
Loop Voltage Limits
1 VDC min., 55 VDC max.
Regulated Loop Power
Regulated Power
Resolution
na
Range
24V dc
Accuracy
± (0.01% of reading + 0.015% FS)
Open Circuit Voltage
24 VDC
Output Drive
15 VDC minimum @ 24 mA, Resistive load
Notes: 1. Nominal resistance at VMA current terminal is 10 – 15 Ω
2. Output load line is linear
Auto-Step Setup Display
Vma Setup
Auto-Step 1:
Start
Mode:
mA Source
Low pt:
4.000 mA
Upr pt:
20.000 mA
Steps:
5
Auto-Step 3
Dwell Time: Manual
Up Down Select Back