41
Save AsFnd to AsLft – accept the first set of saved cal data (As found) to be the As left data. The Pass
indicator on this screen indicates this option is acceptable. The Fail indicator lets the user know if trim
operations and subsequent As left documentation are needed.
Quick Cal – displays HART Pres (PV) and AO along with MFTs measured values for at-a-glance
determination of trim needs. Direct access to the Smart Trim function (see below) is provided through the
Trim soft key. See Quick Cal section of this manual for more details.
Smart Trim – Provides immediate access to HART analog and sensor trim options. See Smart Trim
section of this manual for more details.
As left – Reactivates the From PC (Stored) Procedure Documentation feature to save As left data
View Results – View saved calibration results from As found work. Returns to Cal Options Screen when
finished
Select the As left calibration option from the Cal Options Screen to document the As left condition of the
device under test (after trim operations are complete). The From PC (Stored) Procedure Calibration
Documentation Screen appears again to guide the user through the necessary procedure points as before.
Press the Save soft key after the Sx value matches the Apply value and is stable. Repeat this process for
each Apply value until all procedure test points are complete and the following screen appears.
Store Procedure Ending Screen
Pass / Fail notice – Provides user with notification of the Pass / Fail status of the As left calibration
documentation.
Enter Cal Note – User may enter a calibration note into MFT memory for the procedure just completed;
uses the MFTs alphanumeric entry features.
View Results – View saved calibration results from As left work. Returns to Cal Options Screen when
finished
Press the Done soft key to end the From PC (Stored) Procedure Calibration documentation session.
Cal
As left: DPT-1003 Pass
Enter Cal Note
View Results
Up Down Select Done
Cal
As Fnd: DPT-1003 FAIL
Save As Fnd to AsLft
Smart Trim
As Left
View Results
Quick Cal
Up Down Select Abrt
Calibration Options Screen for