SP4000 Flow Computer

SERVICE & ANALYSIS

6.15
SERVICE &
ANALYSIS
(Continued)

Connect your Ammeter (current meter) to (+) Pin 14 and (-) Pin 16.

Observe the reading on the ammeter. Using the numeric keys, enter
the actual reading (in mA) and press enter.

Display:

ACTUAL? 4.025 mA
ADJ 4mA PIN 14-16

Observe the reading on the ammeter. Using the numeric keys, enter
the actual reading (in mA) and press enter.

Display:

ACTUAL? 20.017 mA
ADJ 20mA PIN 14-16

Connect your Ammeter (current meter) to (+) Pin 15 and (-) Pin 16.

Observe the reading on the ammeter. Using the numeric keys, enter
the actual reading (in mA) and press enter.

Display:

ACTUAL? 4.041 mA
ADJ 4mA PIN 15-16

Observe the reading on the ammeter. Using the numeric keys, enter
the actual reading (in mA) and press enter.

Display:

ACTUAL? 20.006 mA
ADJ 20mA PIN 15-16

Connect your frequency meter to (+) Pin 12 and (-) Pin 13. This
feature is used to check the pulse output. Calibration is not
performed.

Selection:

OFF, 50 Hz, 10 Hz, 1.0 Hz, 0.1 Hz, 0.0 Hz

Display:

OFF
SIMULATION FREQ.

ANALOG OUTPUT 1
CALIBRATION
(Pins 14 & 16)

ADJ

4 mA

(Pins 14 & 16)

ADJ

20 mA

(Pins 14 & 16)

ANALOG OUTPUT 2
CALIBRATION
(Pins 15 & 16)

ADJ

4 mA

(Pins 15 & 16)

ADJ

20 mA

(Pins 15 & 16)

FREQUENCY OUTPUT
SIMULATION
(Pins 12 & 13)