Flow input – Liquid Controls SP4000 User Manual

Page 68

background image

65

SP4000 Flow Computer

2

RATE

1

TOTAL

Enter the maximum possible frequency of a flowmeter with a digital
output. Using the value entered here, the flow computer selects a
suitable limiting frequency for low pass filter to help suppress
interference from higher frequency signals.

Input:

Max. 5 digit number: 10...40000 (Hz):

Display:

40000 Hz
LOW PASS FILTER

LOW PASS
FILTER

DP FACTOR
(Continued)

The DP-Factor (K

DP

) is then computed using one of the following

equations:

Steam:

Liquid:

Gas:

K

DP

DP-Factor

M

Mass flow

Q

Volumetric flow

Q

REF

Corrected volumetric flow

ε

1

Gas expansion factor

T

Operating temperature

T

CAL

Calibration temperature

p

Differential pressure

ρ

Density at flowing conditions

ρ

REF

Reference density

Note:
The computation accuracy can be enhanced by entering up to 16
values for Reynold's Number DP-Factor in a linearization table (see
"LINEARIZATION"). Each DP-Factor can be calculated using the
above procedure. For every calculation, a sizing sheet is required.
The results have to be entered in the linearization table afterwards.