Ez setup – Liquid Controls SP4000 User Manual

Page 37

background image

34

SP4000 Flow Computer

Select the type of fluid appropriate for your application.

Selection:

GENERIC, DRY AIR, NATURAL GAS, NATURAL GAS
(NX19), HUMID GAS, HUMID AIR, HYDROGEN, ARGON,
METHANE, NITROGEN, CARBON DIOXIDE, PROPANE,
OXYGEN, ETHANE, HELIUM

Display:

NITROGEN

FLUID TYPE

Select the flowmeter type used in your application.

Selection:

LINEAR, SQR LAW, SQR LAW-LIN., LINEAR 16 PT, SQR
LAW 16 PT, SQR LAW-LIN. 16 PT, LINEAR UVC, GILFLO,
GILFLO 16 PT, BYPASS

Display:

LINEAR

FLOWMETER TYPE

Select the appropriate input signal.

Selection:

4-20 mA, 0-20 mA, 0-5 Vdc, 1-5 Vdc, 0-10 Vdc, DIGITAL:
10 mV LEVEL, DIGITAL: 100 mV LEVEL, DIGITAL: 2.5 V
LEVEL, 4-20mA STACKED, 0-20mA STACKED, 4-20mA
LINEAR MANIFOLD, 0-20mA LINEAR MANIFOLD

Display:

DIGITAL 2.5 V LEVEL

INPUT SIGNAL

Enter the K-Factor for the flowmeter.

Input:

Number with floating decimal point:
0.0001...999999

Display:

123.67 P/ft3

K-FACTOR

Select the appropriate pressure input signal.

Selection:

MANUAL TEMPERATURE, 4-20 TEMPERATURE,
0-20 TEMPERATURE, RTD TEMPERATURE

Display:

4-20 PRESSURE (ABS.)

INPUT SIGNAL

EZ SETUP

FLUID TYPE

FLOWMETER TYPE

INPUT SIGNAL

K-FACTOR

INPUT SIGNAL
(TEMPERATURE)

2

RATE

1

TOTAL

6.2
EZ
SETUP
(Continued)