39

SP4000 Flow Computer

SYSTEM PARAMETERS

6.4
SYSTEM
PARAMETERS
(Continued)

ENTER TIME

PRIVATE CODE

Special Note:
After returning to the run
mode, program editing
is automatically locked
after 60 seconds as long
as no keys are pressed
The program editing can
also be disabled by
entering a number other
than the private code at
the Access Code
prompt.

SERVICE CODE

Note:
The Service Code will allow
access to the same
information as the Private
Code with the following
additional functions:

• Change the Service Code
• Change the Order Code
• Change the Serial No.
• Clear Grand Total
• Clear Errors in Error Log
• View & Perform calibra-

tion in Service & Analysis
Menu

• Restore Factory Calibra-

tion Information in Service
& Analysis Menu

• Set Next Calibration Date
• Print Maint.Report
• Perform Service Test

Enter the actual time in this format: Hours - Minutes

Note:

After prolonged breaks in the power supply (several days) or
upon initial start-up of the unit, the date and time must be
reset.

Input:

Flashing selections can be changed.
Store and Confirm entries with the ENTER key

Display:

13:24
ENTER TIME

A personal code may be defined. This code is used to enable
program editing.

Note:

• The private code is factory set to 1000
• Entering a private code of "0" will always enable program

editing (Turns automatic lock off)

Input:

Maximum 4 digit number: 0...9999
Store and Confirm entries with the ENTER key

Display:

1000
PRIVATE CODE

A personal service code may be defined. This code is used to enable
program menus that are normally reserved for factory and service
personnel.
(i.e.: Service & Analysis Submenu Group)

Note:

• The service code is factory set to 2000
• The service code submenu will only appear if the service

code was entered for the "Access Code".

Input:

Maximum 4 digit number: 0...9999
Store and Confirm entries with the ENTER key

Display:

2000
SERVICE CODE

ENTER

2

RATE

1

TOTAL

CLEAR

ENTER

2

RATE

1

TOTAL

CLEAR

ENTER

2

RATE

1

TOTAL

CLEAR