SP4000 Flow Computer

SERVICE & ANALYSIS

6.15
SERVICE &
ANALYSIS
(Continued)

This feature allows the calibration of the units inputs and outputs.

CAUTION:

The calibration should only be performed by qualified technicians.
The calibration procedure requires the use of precision Voltage &
Current sources, a frequency generator, a 100

resistor (

±

0.1%),

an ammeter, an ohmmeter and a frequency counter. If calibration
fails, use the "Restore Factory Calibration" feature.

Selection:

NO, YES

Display:

PERFORM? YES
CALIBRATION

Connect your voltage source to (+) Pin 2 and (-) Pin 4.

Apply 0.0 Volts. Press enter to learn 0.0 Volts.

Display:

RESULT: 0.000 V
LEARN 0.0 V PIN 2

Apply 10.0 Volts. Press enter to learn 10.0 Volts.

Display:

RESULT: 10.000 V
LEARN 10.0 V PIN 2

Connect your current source to (+) Pin 2 and (-) Pin 4.

Apply 0.0 mA. Press enter to learn 0.0 mA.

Display:

RESULT: 0.000 mA
LEARN 0.0 mA PIN 2

Apply 20.0 mA. Press enter to learn 20.0 mA.

Display:

RESULT: 20.000 mA
LEARN 20.0 mA PIN 2

Connect your current source to (+) Pin 3 and (-) Pin 4.

Apply 0.0 mA. Press enter to learn 0.0 mA.

Display:

RESULT: 0.000 mA
LEARN 0.0 mA PIN 3

Apply 20.0 mA. Press enter to learn 20.0 mA.

Display:

RESULT: 20.000 mA
LEARN 20.0 mA PIN 3

PERFORM
CALIBRATION

NOTE:
This menu item will only
appear if editing is
enabled with Service
Code.

VOLTAGE INPUT
CALIBRATION

LEARN

0.0 V

(Pin 2)

LEARN

10.0 V

(Pin 2)

CURRENT INPUT
CALIBRATION

LEARN

0.0 mA

(Pin 2)

LEARN

20.0 mA

(Pin 2)

LEARN

0.0 mA

(Pin 3)

LEARN

20.0 mA

(Pin 3)