Connect your current source to (+) Pin 7 and (-) Pin 4.
Apply 0.0 mA. Press enter to learn 0.0 mA.
Display:
RESULT: 0.000 mA
LEARN 0.0 mA PIN 7
Apply 20.0 mA. Press enter to learn 20.0 mA.
Display:
RESULT: 20.000 mA
LEARN 20.0 mA PIN 7
Connect your current source to (+) Pin 11 and (-) Pin 4.
Apply 0.0 mA. Press enter to learn 0.0 mA.
Display:
RESULT: 0.000 mA
LEARN 0.0 mA PIN 11
Apply 20.0 mA. Press enter to learn 20.0 mA.
Display:
RESULT: 20.000 mA
LEARN 20.0 mA PIN 11
Connect a 100
Ω
resistor between Pins 6 & 7 and place a jumper
wire between Pins 5 & 6.
Press enter to learn RTD resistance on Pins 5, 6 & 7.
Display:
RESULT: 100.00 ohm
LEARN RTD PIN 5-6-7
Connect a 100
Ω
resistor between Pins 10 & 11 and place a jumper
wire between Pins 9 & 10.
Press enter to learn RTD resistance on Pins 9, 10 & 11.
Display:
RESULT: 100.00 ohm
LEARN RTD PIN 9-10-11