beautypg.com

Service & analysis – Liquid Controls SP4000 User Manual

Page 96

background image

93

SP4000 Flow Computer

SERVICE & ANALYSIS

6.15
SERVICE &
ANALYSIS
(Continued)

Connect your current source to (+) Pin 7 and (-) Pin 4.

Apply 0.0 mA. Press enter to learn 0.0 mA.

Display:

RESULT: 0.000 mA
LEARN 0.0 mA PIN 7

Apply 20.0 mA. Press enter to learn 20.0 mA.

Display:

RESULT: 20.000 mA
LEARN 20.0 mA PIN 7

Connect your current source to (+) Pin 11 and (-) Pin 4.

Apply 0.0 mA. Press enter to learn 0.0 mA.

Display:

RESULT: 0.000 mA
LEARN 0.0 mA PIN 11

Apply 20.0 mA. Press enter to learn 20.0 mA.

Display:

RESULT: 20.000 mA
LEARN 20.0 mA PIN 11

Connect a 100

resistor between Pins 6 & 7 and place a jumper

wire between Pins 5 & 6.

Press enter to learn RTD resistance on Pins 5, 6 & 7.

Display:

RESULT: 100.00 ohm
LEARN RTD PIN 5-6-7

Connect a 100

resistor between Pins 10 & 11 and place a jumper

wire between Pins 9 & 10.

Press enter to learn RTD resistance on Pins 9, 10 & 11.

Display:

RESULT: 100.00 ohm
LEARN RTD PIN 9-10-11

CURRENT INPUT
CALIBRATION
(continued)

LEARN

0.0 mA

(Pin 7)

LEARN

20.0 mA

(Pin 7)

LEARN

0.0 mA

(Pin 11)

LEARN

20.0 mA

(Pin 11)

RTD INPUT
CALIBRATION

Temperature

Input

(Pins 5, 6 & 7)

Temperature 2

Input

(Pins 9, 10 & 11)