Screen 19c: Preset and Delivery (3/7)
Preset and Delivery
The third Preset and Delivery screen is used to set a few
delivery options.
Product # and Units will reflect values
selected in the previous screens.
Product Type is the current product type associated
with the selected
Product # as defined in the Calibration
screens.
Aux1 and Aux2 are used to define the function of auxiliary
outputs 1 and 2. If there are no external devices attached
to these two outputs, there is no need set these values.
Move the pointer to
Aux1 and press ENTER to open a
list box window with the following options:
Monitor Flow Rate: This option causes auxiliary 1 to
activate at the beginning of each delivery. If the flow
rate during the delivery meets or exceeds 40 units
per minute, auxiliary 1 will deactivate. This option is
typically used to monitor the amount of restriction in
a strainer basket or filter. The idea being, when the
filter or strainer is getting clogged, the flow rate will
decrease.
Off: This option deactivates auxiliary 1.
On: This option activates auxiliary 1. It remains activated
even when a delivery is not taking place.
On During Delivery: This option activates auxiliary output
only when the system is in an active delivery. This
also causes header line 11 to print on the ticket.
Move the pointer to
Aux2 and press ENTER to open a
list box windows with the following options:
Flow Direction: This option causes auxiliary 2 to indicate
the flow direction through the flow meter.
Off, On, On During Delivery are all defined as in Aux1
above. If set to
On During Delivery, header line 12
will print on the ticket.
No-Flow Timer is used to set the no flow timer for the
next delivery. Move the pointer to
No-Flow Timer and
press
ENTER to open a field edit window. Enter a value
between “
0” and 3600 seconds. This timer will end an
active delivery if there is no flow through the meter in
the set amount of time. If this value is set to “
0”, then
each delivery ticket will contain the text
“MULTIPLE
DELIVERIES AT ONE SITE. This value is typically set
to
180 (3 minutes).
Press
Next (F1) to advance to Screen 19d on page 57.