Setup network – Liquid Controls DMS Setup User Manual
Page 22
22
Screen 4b: Setup Network (2/3)
Setup Network
(2/3)
1|1-METER SYSTEM
1
METER-1
SETUP NETWORK
Network:
Device:
Name:
Next Prev New Del
(2/3)
1|1-METER SYSTEM
SETUP NETWORK
Network:
New Device Number
|1_ |
Home <-- Ins --> End
(2/3)
1|1-METER SYSTEM
SETUP NETWORK
Network:
Device Name
|METER-1 |
Home <-- Ins --> End
(2/3)
1|1-METER SYSTEM
1
METER-1
SETUP NETWORK
Network:
Device:
Name:
Next Prev New Del
This screen is used to establish communication with
LectroCount Registers (LCRs). By default, all LCRs are
shipped with a node address of 250. This information
will be helpful when establishing communication with
new LCRs.
Based on examples provided, one network has been
defined and this is
Network # 1. Move pointer to Device
and select
New (F3) The device number can be between
1 and 249. Enter the number “1” and press ENTER to
accept.
Next, move pointer to
Name and press New (F3) to open
a field edit window. This field can be from 0 to 16 alpha-
numeric characters in length. Enter a suitable name for
the device number 1. For example, type in “
LCR1” and
press
ENTER to accept.