Save configuration, Screen 14: save configuration – Liquid Controls DMS Setup User Manual
Page 51

51
Save Configuration
Screen 14: Save Configuration
This feature allows the user of LCRHost to save
the calibration and configuration parameters of an
LectroCount in a file. This file can then be used to set up
other LectroCount registers with same data without having
to manually enter data individually.
Press
ENTER to open a field edit window and enter an
alphanumeric
Output File name up to 15 characters in
length. When complete, press
Save (F1) to save output
file. When
Save (F1) is pressed, system will prompt the
operator with the list box window that asks if the current
file should be overwritten with the selected
Output File.
Select
Yes or No and press ENTER.
The fields saved in the output file are:
DateFormat
Decimals
FlowDirection
MeterID
NoFlowTimer
Odometer
PresetsAllowed
PresetType
Printer
PrintGrossAndParam
PulseOutputEdge
PulsesPerDistance
QtyUnits
Residual
SaleNumber
TempScale
TicketNumber
TicketRequired
TimeUnit
UnitID
VolCorrecteDMS i1000g
For each of the four
product calibrations:
Aux1
Aux2
AuxMult
AuxUnit
BaseTemp
CompensationParam
CompensationType
GrossPreset
GrossTotal
Linearize
NetPreset
NetTotal
PercentTax
PricePerUnit
ProductCode
ProductDescriptor
ProductType
PulsesPerUnit
S1Close
TaxPerUnit
For each of the 10
linearization points:
LinearFlowRate
LinearError
For each of the 12 ticket
header lines:
TicketHeaderLine
Press
Main (F5) to return to Screen 2 on page 20. Move
pointer to
Load Configuration and press ENTER to
advance display to
Screen 18 on page 52.
1
SAVE CONFIGURATION
Use the field below to specify
the output file that will
receive the configuration data
from the LCR.
Output File:
Save LCR# Main