beautypg.com

System calibration, Screen 9a: system calibration (1/5) – Liquid Controls DMS Setup User Manual

Page 32

background image

32

Screen 9a: System Calibration (1/5)

System Calibration screens cover configuration and

calibration of the metering system.
Meter ID is an alpha-numeric value used to uniquely

identify a Meter attached to a LectroCount register. It is

important a unique number is selected and recommended

the meter serial number be entered. This number will

print on all calibration and diagnostic tickets for future

reference.
Press

ENTER to open a field edit window and enter a

unique identifier up to 10 alpha-numeric characters in

length. Press

ENTER to accept this value.

Flow Direction is used to set direction of flow through

the meter. If the LectroCount display counts in reverse

(decreasing numbers) after unit is installed, it will be

necessary to select opposite

Flow Direction.

Ticket Required is used to select whether a delivery

ticket will be required for each delivery. Most Weights

& Measures governed truck applications will require a

ticket.
There are two options for

Ticket Required?. If No is

selected, the system will

NOT require a ticket to be printed

after each delivery. If

Yes is selected, the LectroCount

will not allow deliveries to be made unless a ticket is in the

printer and the printer is operational. It will also require a

ticket be printed before another delivery can be started.
Move pointer to desired option and press

ENTER to

accept.
NOTE: Selecting No will still allow a ticket to be printed

if a printer is connected, loaded with paper, and ready

to print.
Printer is used to select printer model to communicate

with the LectroCount. Selecting a wrong printer model

may result in miscommunication between LectroCount

and printer.
There are six options for the

Printer. Use the following

guide to select the correct option:
BLASTER (select for use with Cognitive Solutions

Thermal Printer).

EPSON NewFontA (select for use with EPSON TM-

T88iii)

EPSON NewFontB (select for use with EPSON 200 Roll

and EPSON 220 Roll)

EPSON OldFontA (select for use with EPSON 290 Slip

and EPSON 295 Slip)

EPSON OldFontB (select for use with EPSON 300

Roll)

Okidata ML184T (select for use with Okidata ML184T)

System Calibration

dP Shutdown Value represents differential pressure

that will cause current active delivery to shutdown. This

function is only active when the following conditions

exist:
1. A Differential Pressure device is installed.

2. LectroCount Register has

SR214 software installed.

3. Value is set to a number other than “

0

When these three conditions exist, reaching or exceeding

the

dP Shutdown Value will end the delivery. Press

ENTER to open a field edit window and enter a value

between

0 and 59.9.

Press

Next (F1) to advance to Screen 9b on page 33.

(1/5) 1

4545

->

Yes

Epson OldFontA

xx.x

SYSTEM CALIBRATION

Meter ID:

Flow Direction:
Ticket Required?

Printer

dP Shutdown Value:

Next Cnfg Main