Burkert Type 8136 User Manual
Page 51

Code
Text mes-
sage
Cause
Rectification
F125
Unpermissible
electronics
temperature
l Temperature of the elec-
tronics in the non-specified
section
l Check ambient tempera-
ture
l Isolate electronics
l Use instrument with higher
temperature range
F260
Error in the
calibration
l Error in the calibration car-
ried out in the factory
l Error in the EEPROM
l Exchange of the elec-
tronics
l Send instrument for repair
F261
Error in the
configuration
l Error during setup
l False signal suppression
faulty
l Error when carrying out a
reset
l Repeat setup
l Repeat reset
F264
Installation/S-
etup error
l Adjustment not within the
vessel height/measuring
range
l Max. measuring range of
the instrument not suffi-
cient
l Check or correct installa-
tion and/or parameter ad-
justment
l Use an instrument with
bigger measuring range
F265
Measurement
function dis-
turbed
l Sensor does no longer
carry out a measurement
l Operating voltage too low
l Check operating voltage
l Carry out a reset
l Separate operating voltage
briefly
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
C700
Simulation
l Simulation active
l Finish simulation
l Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Function check
Out of specification
LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire
51
9 Diagnosis, Asset Management and service
41783
-EN
-
120316