T250 ma, English, 34 installationandwiring – Burkert Type 8035 User Manual
Page 36
34
Installationandwiring
230V
L N
230V
T
250 mA
5 6 7 8 9 10
L N
230V
1
A
2
4
B
C
CURRENT
SOURCE
SINK
BINARY
PE
PE
PE
PE
DI1
DI2
DI3
DI4
DO4
ISOG
FLOW
SENSOR
L+
L-
PE P- P+
NC
Iout
PULSE
DO1
Supply
12..36Vdc
Univ
Batch
(AO1)
SUPPLY
NC
COIL
PULSE INPUT
NPN/PN
P
2
1
3
PE
+
-
SENSOR
SUPPL
Y
LOAD
+5V
L+
(L+)-12V
COIL/PNP
39K
470
2.2K
SENSOR TYPE
COIL
NPN/PNP
DO2
DO3
OFF
ON
5
6
D
7
8
9
3
terminal block 1
PE: factory wired shield.
terminal block 2
• NC: not connected
• L+: V+ (red wire, factory wired)
• L-: 0V (black wire, factory wired)
• PE: protective earth, factory wired
• P-: negative transistor output (DO1) (brown wire, factory
wired)
• P+: positive transistor output (DO1) (white wire, factory
wired)
terminal block 3 "binAry"
• DI1 to DI4: 4 digital inputs
• DO4: transistor output
• ISOG: ISOGND, ground common to the 4 digital inputs
and the transistor output DO4.
• PE: cable shielding
terminal block 4 "floW sensor": Wiring the remote
flow sensor. The wiring depends on the type of output
signal originating from the flow sensor: see chap. “7.6.7”.
terminal block 5: wiring the DO2 relay output.
terminal block 6: wiring the DO3 relay output.
terminal block 7
• terminal 5 not connected
• terminal 6: positive 27 V DC power supply,
available to energize an external instrument
• terminal 7: 0V (earth of the power supply
available to energize an external instrument)
• terminal 8: protective earth for the cable
shieldings
• terminal 9: negative transistor output (DO1)
• terminal 10: positive transistor output (DO1)
terminal block 8: wiring of the 115/230 V AC
power supply
mark 9 : time-delay fuse to protect the
115/230 V AC power supply
selector A : see chap. “7.6.7”.
selector B : see “Table 5”, page 40.
selector C : see chap. “7.6.7”.
selector D : see chap. “7.6.7”.
Figure 21:
Terminal assignment of a wall-mounted version, 115/230 V AC
English
Type 8025 - 8035 BATCH