beautypg.com

Pipe preparation and transducer mounting – Dynasonics DXN Portable Ultrasonic Measurement System User Manual

Page 9

background image

PIPE PREPARATION AND TRANSDUCER MOUNTING

DTTN and DTTL Transit Time Transducers

1. During this procedure, observe the flowmeter’s signal quality value. This value is available on the DXN Display>Meters

where TT Quality is one of the parameters that can be shown. Signal quality value should be at least 10%.

2. The pipe surface where the transducers are to be mounted must be clean and dry. Remove scale, rust or loose paint

to ensure satisfactory acoustic conduction. Wire brushing the rough surfaces of pipes to smooth bare metal may also

be useful. Plastic pipes do not require preparation other than cleaning.

3. Apply a single 1/2" (12 mm) wide by approximately 1/8" thick bead of acoustic couplant grease to the upstream

transducer and secure it to the pipe with a mounting strap.

4. Apply acoustic couplant grease, as in step 3 above, to the downstream transducer and press it onto the pipe using

hand pressure at the calculated lineal distance.

5. Space the transducers according to the recommended values found during programming. Secure the transducers

with the mounting straps at these locations while making sure to use the alignment marks on the transducers for

proper spacing.

DTTSU Universal Small Pipe Transit Time Transducers

1. During this procedure, observe the flowmeter’s signal quality value. This value is available in Display>Meters where TT

Quality is one of the parameters that can be shown. Signal quality value should be at least 10%.

2. The pipe surface where the transducers are to be mounted must be clean and dry. Remove scale, rust or loose paint

to ensure satisfactory acoustic conduction. Wire brushing the rough surfaces of pipes to smooth bare metal may also

be useful. Plastic pipes do not require preparation other than cleaning.

3. Set the downstream transducer spacing to the value found on the Site>Transit.
4. Apply a single 1/2" (12 mm) wide by approximately 1/8" thick bead of acoustic couplant grease to the face of each

transducer and secure the unit to the pipe with attachment chains.

5. Tighten the two thumb screws located on the top of the unit evenly so that the acoustic coupling grease begins

to flow out from the edges of the transducer and from the gap between the transducer and the pipe. Do not

over tighten.

DT94 Doppler Transducers

1. In general, select a mounting location on the piping system with a minimum of 10 pipe diameters (10 ×

the pipe inside diameter) of straight pipe upstream and 5 straight diameters downstream. See

Table 1

for

additional configurations.

2. The pipe surface, where the transducers are to be mounted, must be clean and dry. Remove scale, rust or loose paint

to ensure satisfactory acoustic conduction. Wire brushing the rough surfaces of pipes to smooth bare metal may also

be useful. Plastic pipes do not require preparation other than cleaning.

3. Apply a single 1/2" (12 mm) wide by approximately 1/8" thick bead of acoustic couplant grease to both transducers

and secure them to the pipe 180° apart using a mounting strap. Ensure that the transducer cable is pointing in the

downstream direction.

4. Record the value calculated and displayed as "Required Spacing" in the Site>Transit.

Installation & Operation Manual

Page 9

February 2013