beautypg.com

Test outputs (test out) – Dynasonics DXN Portable Ultrasonic Measurement System User Manual

Page 67

background image

The I/O>Scale Out sets the parameter the output circuitry will respond to. Choices for Data To Output are either Flow or Power.

Additionally this screen sets the minimum and maximum values the outputs will be scaled to.

NOTE:

N

Transit time measurements are capable of bidirectional flow but Doppler is not. The only time setting the Flow at Min

Out to a negative value may be necessary is if the meter stays in transit time mode. If the meter is used in Doppler

mode exclusively or when in hybrid mode the meter switches to Doppler mode, set the value for Flow at Min Out no

lower than zero.

An example of a valid use of setting Flow at Min Out below zero would be a transit time application where flow can be in

either the forward or reverse direction. If for instance a tap water system is capable of 100 gpm forward and 100 gpm in

reverse then setting the Flow at Min Out to -100 and Flow at Max Out to +100 would be valid entries.
If the meter were programed to output a 4-20 mA signal then 4 mA would represent –100 gpm and 20 mA would indicate

+100 gpm. The zero flow point would be indicated as 12 mA (halfway between 4 mA and 20 mA).

-100

100

-50

50

0

GPM

4 mA

20 mA

8 mA

16 ma

12 mA

mA

Dead

Band

Figure 86: Bidirectional current output

Also keep in mind that no flowmeter can read down to zero. There is going to be a small dead band around true zero where

the velocity of the fluid is not great enough for the meter to register. The absolute minimum under perfect conditions is 0.1

fps (0.03 mps). Whatever flow rates these velocities correspond to will be the minimum dead band around true zero.

Test Outputs (Test Out)

Site Name

0.00 GPM

3/28/2012

13:24

I/O

Display

Adv

Cal

System

Site

Meter

Log

Test Out

Test Frequency Out (Hz)

20

Test 4-20 mA Out (mA)

4.000

Test 0 - 10V Out (V)

0.000

Enter Output Test Mode

Scale Out

Set Out

Figure 87: Test Outputs Setup

Test Out is used to calibrate devices connected to the DXN thru the I/O breakout box. To use this function first connect the

desired output to a device designed to read that type of output signal. For 4-20 mA out use a milliammeter. For the 0...10V DC

output use a voltmeter and for the frequency output either a frequency counter or oscilloscope is necessary.
Next using the + and - buttons select the output level to calibrate to. For example common test levels for the 4-20 mA output

are 4, 8, 12, 16, and 20 mA.
Finally put a check mark in the Enter Output Test Mode box to activate the outputs.

Installation & Operation Manual

Page 67

February 2013