In

Figure 76

flow data falls outside the flow hysteresis window but does not reach the minimum time specified in the bad

data window. When data appears that is outside the hysteresis band and shorter than the bad data window time the data is

rejected.

±10% Hysteresis

Limit

Flow Outside

Hysteresis Limit

3 Samples Outside

Hysteresis Limit

Bad Data Rejection

Window

Flow

1

2 3

Flow Outside

Hysteresis Limit

3 Samples Outside

Hysteresis Limit

±10% Hysteresis

Limit

Bad Data Rejection

Window

Flow

1

2 3

150

100

50

0

110

90

Sample

Limits

Figure 76: Bad Data (Rejection)

The flow rate is again outside the original ±10% hysteresis window but the data exists for a time period greater than the Bad

Data Window. In this instance the meter interprets the data as a new valid flow rate and moves the hysteresis window to

correspond with the new established flow rate.