beautypg.com

Dynasonics DXN Portable Ultrasonic Measurement System User Manual

Page 22

background image

Transducer Mount Mode

Transducer

Pipe Material

Pipe Size

W-Mount

DTTSU

Plastic (all types)

0.5…0.75" (12.7…19 mm)

Carbon Steel

Stainless Steel

Copper

Galvanized

Not recommended

V-Mount

Plastic (all types)

0.75…2.4" (19…61 mm)

Carbon Steel

Stainless Steel

Copper

Galvanized

0.5…2.4" (12.7…61 mm)

Table 3: Starting Point Transducer Mounting Modes—DTTSU

For pipes 24" (600 mm) and larger the DTTL transducers using a transmission frequency of 500 KHz are recommended.
DTTL transducers may also be advantageous on pipes 4…24" if there are less quantifiable complicating aspects such as –

sludge, tuberculation, scale, rubber liners, plastic liners, thick mortar, gas bubbles, suspended solids, emulsions, or pipes that

are perhaps partially buried where a V-mount is required/desired, etc.
The DTTN, DTTH and DTTL transducers must be properly oriented and spaced on the pipe to provide optimum reliability and

performance. On horizontal pipes, when Z-Mount is required, the transducers should be mounted 180 radial degrees from

one another and at least 45 degrees from the top-dead-center and bottom-dead-center of the pipe. See

Figure 24

. Also see

Z-Mount Configuration

. On vertical pipes the orientation is not critical.

The best accuracy is achieved when transducer spacing is exactly what the DXN calculates, so the calculated spacing should

be used if signal quality is satisfactory. If the pipe is not round, the wall thickness not correct or the actual liquid being

measured has a different sound speed than the liquid programmed into the transmitter, the spacing can vary from the

calculated value. If that is the case, the transducers should be placed at the highest signal level observed by moving the

transducers slowly around the mount area.

NOTE:

N

Transducer spacing is calculated on ideal pipe. Ideal pipe is almost never found so the transducer spacing distances

may need to be altered. An effective way to maximize signal quality is to configure the display to show signal quality,

fix one transducer on the pipe and then starting at the calculated spacing, move the remaining transducer small

distances forward and back to find the maximum signal quality point.

45°

45°

YES

YES

45°

45°

TOP OF

PIPE

Figure 24: Transducer Orientation – Horizontal Pipes

The spacing between the transducers is measured between the two spacing marks on the sides of the transducers. These

marks are approximately 0.75" (19 mm) back from the nose of the DTTN and DTTH transducers, and 1.2" (30 mm) back from

the nose of the DTTL transducers. See

Figure 25

.

DXN Portable Ultrasonic Measurement System

Page 22

February 2013