beautypg.com

Dynasonics DXN Portable Ultrasonic Measurement System User Manual

Page 44

background image

Pipe Material

Choose the pipe material from the combo box drop-down.

Manual Dimension Entry Iron - Cast

St Steel 304L

Aluminum

Iron - Ductile

St Steel 316

Brass (Naval)

PVC CPVC

St Steel 347

Carbon Steel

St Steel 302/303

St Steel 410

Copper

St Steel 304

St Steel 430

Table 9: Pipe Material Choices

NOTE:

N

This list is provided as an example. Additional pipe materials are added periodically.

Pipe Schedule/Class

The choice of pipe material determines the choices available in the Schedule/Class drop-down menu. For example, if a pipe

material that is governed by ANSI standards is selected, the menu label will show Schedule and the schedule choices will be

appropriate for ANSI pipe. If Cast Iron is chosen the label Class will be shown instead of Schedule, indicating the pipe material

is categorized in classes.

Nominal Size

This selection allows various standard or nominal pipe sizes to be chosen. Simply choose one of the nominal values from the

drop-down menu.

Liner Schedule – Restricted Choice

Certain combinations of Pipe Material, Class, and Nominal Size have liner information built into the pipe specification. This

condition is usually found in ductile iron classes, 6 inches and up. If a ductile pipe with built-in liner is used, a selection of

either Standard or Double can be made.

NOTE:

N

The choice of Standard or Double will disable the Site>Liner menu item.

Site Name

0.00 GPM

1/16/2013

2:18 PM

Meter

Adv

Cal

System

Display

Site

Log

I/O

Create

Liner

Transit

Doppler

Fluid

Pipe

Lookup

Pipe Material

Manual Dimension Entry

Schedule

Nominal Size (in)

Liner Type

Use Manual Entry

Use Manual Entry

Use Manual Entry

Figure 57: Lookup Page Manual Entry

If Manual Dimension Entry is selected in Site>Lookup (see

Figure 57

) screen the parameters in the Site>Pipe screen will be

available for user input as in

Figure 59

.

NOTE:

N

If the Site>Lookup feature is used, the values are automatically loaded into the parameter choices in the Site>Pipe

screen but Pipe OD and Pipe Wall [Thickness] are still available for changes.

DXN Portable Ultrasonic Measurement System

Page 44

February 2013