CALIBRATION (CAL) GROUP

Meter

Adv

Display

System

Site

Log

I/O

Cal

Factory

Transit

Doppler

Factory

Transit

Doppler

Figure 95: Advanced Group

Transit

Site Name

0.00 GPM

1/16/2013

2:18 PM

Cal

Meter

Adv

Display

System

Site

Log

I/O

Transit

Factory

Doppler

User Adjustment

1.000

0.000

0.000

Positive

Flow

Reynolds

Correcting

SOS

Comp

Soundspeed Offset (us)

Delta Time Zero (ns)

Figure 96: Calibrate Transit Time

Press Delta Time Zero (ns) zeros to remove the No Flow transit time offset. This is also referred to as Zeroing the meter.
Because every flowmeter installation is slightly different and sound waves can travel in slightly different ways through these

various installations, it is important to remove the zero offset at zero flow to maintain the meter’s accuracy. A provision is

made using this button to establish Zero flow and eliminate the offset.

Procedure:

1. The pipe must be full of liquid.
2. Flow must be absolutely zero. Securely close any valves and allow time for any settling to occur.
3. Press the Transit Time Zero button (Delta Time Zero) once.