beautypg.com

Transducer installation, General guidelines – Dynasonics DXN Portable Ultrasonic Measurement System User Manual

Page 18

background image

Water-Tap

Set

×

Water-Tap

Water-Black

Acetone

Ammonia

Benzine

Brine

Ethanol

Gasoline

Glycerin

Isopropanol

Kerosen

Methanol

Milk 4%

Oil Diesel

Oil Hydraulic (Petro)

Oil Lubricating

Parameter

Choices

Move

Selection

Buttons

Make

Selection

Scroll

Bar

Cancel

Figure 21: Combo Box with Large Scroll Bars

The

and

arrow buttons moves the blue highlights area up or down depending on the original position of the

highlight. Once the correct choice has been highlighted in blue use the set

Set

button to lock in the selection. If no change is

desired use the cancel

×

button to exit the combo box without making any changes.

If the drop-down list of parameters is too large to be contained in one combo box length a large scroll bar will be visible

allowing the list to be scrolled up or down.
A small scroll bar is also used when parameter entries are too numerous for a single screen page. The scroll bars may appear

either horizontally or vertically.

Figure 22: Small Scroll Bar

TRANSDUCER INSTALLATION

General Guidelines

At its simplest setting up the DXN portable ultrasonic measurement system consists of two steps. The first step is configuring

the electronics. The second step, mounting the transducers, requires information obtained when the electronics are

configured.

Transmission Mode

The first decision to be made is what mode of ultrasonic transmission should be used. Generally, the transit time mode is used

with fluids that have little to no particulates or entrained air. Doppler mode is used on fluids that contain significant levels of

particulates and aerated fluids. If the fluid type is unknown it is usually better to start with transit time as transit time mode is

more forgiving of having some particulate or air than Doppler is of not having enough particulate or entrained air.
If the fluid type is an unknown it is also possible to mount both transit time and Doppler transducers on the same pipe and let

the DXN decide which set of transducers to use. The DXN can operate as a dedicated transit time meter, a dedicated Doppler

meter or to automatically switch between transit time and Doppler modes using the Hybrid mode.

Transducer Mounting Location

At this point it is also necessary to consider the transducer mounting location. A very loose rule of thumb is to mount the

transducers on a section of pipe that has at least 10 pipe diameters upstream of the transducers and 5 pipe diameters

downstream. See

Table 1

for additional pipe length considerations.

For example, if a 3" pipe is being measured the minimum upstream pipe in front of the transducers should be 30" and the

minimum downstream pipe behind the transducers should be at least 15".

DXN Portable Ultrasonic Measurement System

Page 18

February 2013