beautypg.com

Hale 5.0 FoamLogix User Manual

Page 87

background image

ΙΙΙ

-7

ROTARY GEAR PUMP

ELECTRONIC FOAM PROPORTIONING SYSTEM

Section

III

: Set-up and Calibration

WATER FLOWSENSOR CALIBRATION

NOTE: The flowsensor is calibrated at
Hale Products Inc. and matched to the
control unit. If the system is installed
properly only minor adjustments should
be necessary to flowsensor reading.

Flowsensor calibration should be verified
during NFPA/UL testing of apparatus and
delivery to end user.

NOTE: An accurate flow measuring
device must be used to measure the
water flow when calibrating the
flowsensor. Use a suitable size smooth
bore nozzle and an accurate Pitot
Gauge instrument. Hand held pitot
gauges are usually not very accurate.
Make sure the system is calibrated with
an accurate flow measuring device.

Determine the water flow normally
expected from the discharge outlet and
establish flow. Make sure the water flow
established is within the range of the
flowsensor monitoring the discharge. For
example, actually establish a flow of 150
GPM (568 LPM) of water through a nozzle
and Pitot system. Compare the calculated
flow value to the value shown on the
control unit digital display.

Press the

é or ê button to set the reading

to match the actual flow calculated from
the Pitot gauge reading. Decrease fire
pump pressure by approximately ½ and
recalculate water flow rate. Verify the
reading on the control unit is within 5% of
the actual value. Stop the water flow when
the reading adjustments are completed.

Figure 3-4. Flowsensor Calibration Display

Figure 3-3. Entering Calibration Mode

RELEASE WHEN

DISPLAY SHOWS

"CALIB"

PRESS AND

HOLD

CONTINUE TO

HOLD

DISPLAYS WHEN

ê

IS PRESSED

DISPLAYS WHEN

é

IS PRESSED

PRESS IN

SEQUENCE AS

NECESSARY TO

ENTER

PASSWORD

This manual is related to the following products: