2 configuration procedure, Configuration procedure – Emerson Process Management MICRO MOTION 2400S User Manual
Page 85
![background image](/manuals/82275/85/background.png)
Configuration and Use Manual
77
Optional Configuration
Requ
ired Co
nfi
g
u
ratio
n
Op
ti
on
al
Co
nfi
g
u
rati
o
n
Using the T
rans
m
itte
r
Us
ing a De
vice
Net T
o
ol
Requ
ired Co
nfi
g
u
ratio
n
Op
ti
on
al
Co
nfi
g
u
rati
o
n
Using the T
rans
m
itte
r
Us
ing a De
vice
Net T
o
ol
Requ
ired Co
nfi
g
u
ratio
n
Op
ti
on
al
Co
nfi
g
u
rati
o
n
Using the T
rans
m
itte
r
Us
ing a De
vice
Net T
o
ol
Requ
ired Co
nfi
g
u
ratio
n
Op
ti
on
al
Co
nfi
g
u
rati
o
n
Using the T
rans
m
itte
r
Us
ing a De
vice
Net T
o
ol
Table 8-14 summarizes these options.
8.13.2
Configuration procedure
The API configuration parameters are listed and defined in Table 8-15.
To configure the petroleum measurement application:
•
Using ProLink II, see Figure B-3.
•
Using a DeviceNet tool, see Table C-10.
Note: This functionality is not available via the display menus.
Table 8-14
API reference temperature tables
Table
CTL
derivation
method
Base temperature
Density unit and range
Degrees API
Base density
Relative density
5A
Method 1
60 °F, non-configurable
0 to +100
5B
Method 1
60 °F, non-configurable
0 to +85
5D
Method 1
60 °F, non-configurable
–10 to +40
23A
Method 1
60 °F, non-configurable
0.6110 to 1.0760
23B
Method 1
60 °F, non-configurable
0.6535 to 1.0760
23D
Method 1
60 °F, non-configurable
0.8520 to 1.1640
53A
Method 1
15 °C, configurable
610 to 1075 kg/m
3
53B
Method 1
15 °C, configurable
653 to 1075 kg/m
3
53D
Method 1
15 °C, configurable
825 to 1164 kg/m
3
Reference temperature
Supports
6C
Method 2
60 °F, non-configurable
60 °F
Degrees API
24C
Method 2
60 °F, non-configurable
60 °F
Relative density
54C
Method 2
15 °C, configurable
15 °C
Base density in kg/m
3
Table 8-15
API parameters
Variable
Description
Table type
Specifies the table that will be used for reference temperature and reference density unit. Select
the table that matches your requirements. See API reference tables.
User defined TEC
(1)
(1) Configurable if Table Type is set to 6C, 24C, or 54C.
Thermal expansion coefficient. Enter the value to be used in CTL calculation.
Temperature units
(2)
(2) In most cases, the temperature unit used by the API reference table should also be the temperature unit configured for the transmitter
to use in general processing. To configure the temperature unit, see Section 6.3.
Read-only. Displays the unit used for reference temperature in the reference table.
Density units
Read-only. Displays the unit used for reference density in the reference table.
Reference
temperature
Read-only unless Table Type is set to 53x or 54x. If configurable:
• Specify the reference temperature to be used in CTL calculation.
• Enter reference temperature in °C.