Emerson Process Management MICRO MOTION 2400S User Manual
Page 84

76
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Optional Configuration
API reference tables
Reference tables are organized by reference temperature, CTL derivation method, liquid type, and
density unit. The table selected here controls all the remaining options.
•
Reference temperature:
-
If you specify a 5x, 6x, 23x, or 24x table, the default reference temperature is 60
°
F, and
cannot be changed.
-
If you specify a 53x or 54x table, the default reference temperature is 15
°
C. However, you
can change the reference temperature, as recommended in some locations (for example, to
14.0 or 14.5
°
C).
•
CTL derivation method:
-
If you specify an odd-numbered table (5, 23, or 53), CTL will be derived using method 1
described above.
-
If you specify an even-numbered table (6, 24, or 54), CTL will be derived using method 2
described above.
•
The letters A, B, C, or D that are used to terminate table names define the type of liquid that the
table is designed for:
-
A tables are used with generalized crude and JP4 applications.
-
B tables are used with generalized products.
-
C tables are used with liquids with a constant base density or known thermal expansion
coefficient.
-
D tables are used with lubricating oils.
•
Different tables use different density units:
-
Degrees API
-
Relative density (SG)
-
Base density (kg/m
3
)