Ii, see figure 10-6, Figure 10-6 prolink ii – flowmeter zero procedure – Emerson Process Management MICRO MOTION 2400S User Manual
Page 110
![background image](/manuals/82275/110/background.png)
102
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Measurement Performance
Figure 10-6
ProLink II – Flowmeter zero procedure
Modify zero time
if required
Calibration
Failure LED
Calibration in Progress
LED turns red
Green
Troubleshoot
Red
Perform Auto Zero
Done
ProLink >
Calibration >
Zero Calibration
Wait until Calibration in
Progress LED turns green