Table 10-1
DeviceNet interface for meter verification
Step number
Step description
Interface
1
Set output state
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 58
Data type: USINT
Value:
• 0: Last measured value (default)
• 1: Fault
Service: Set
2
Set uncertainty limit
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 59
Data type: REAL
Range: 0.0025 to 0.05
Default: 0.04
Service: Set
3
Start/abort procedure
Class: Diagnostics Object (0x66)
Instance: 1
• 0: Abort
• 1: Start
Service: 0x50
4
Check current algorithm state
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 54
Data type: USINT
Service: Get
5
Read percent complete
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 57
Data type: USINT
Service: Get
6
Check algorithm abort state
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 56
Data type: USINT
Service: Get
7
Check inlet stiffness
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 61
Data type: USINT
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Service: Get
8
Check outlet stiffness
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 62
Data type: USINT
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Service: Get
9
Read abort code
Class: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 55
Data type: USINT
Codes: See Table C-7
Service: Get