Table 8-8 includes info – Emerson Process Management MICRO MOTION 2400S User Manual
Page 74

66
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Optional Configuration
Some alarms can be reclassified. For example:
•
The default severity level for Alarm A020 (calibration factors unentered) is
Fault
, but you can
reconfigure it to either
Informational
or
Ignore
.
•
The default severity level for Alarm A102 (drive over-range) is
Informational
, but you can
reconfigure it to either
Ignore
or
Fault
.
For a list of all status alarms and default severity levels, see Table 8-8. (For more information on
status alarms, including possible causes and troubleshooting suggestions, see Table 11-2.)
To configure alarm severity:
•
Using ProLink II, see Figure B-3.
•
Using a DeviceNet tool, refer to Table C-7 and:
a.
Set the alarm index (Attribute 18).
b.
Set the severity for that alarm (Attribute 19).
Note: This functionality is not available via the display menus.
Table 8-8
Status alarms and severity levels
Alarm code
ProLink II message
Default
severity
Configurable?
Affected by
fault timeout?
A001
(E)EPROM Checksum Error (CP)
Fault
No
No
A002
RAM Error (CP)
Fault
No
No
A003
Sensor Failure
Fault
Yes
Yes
A004
Temperature Sensor Failure
Fault
No
Yes
A005
Input Overrange
Fault
Yes
Yes
A006
Not Configured
Fault
Yes
No
A008
Density Overrange
Fault
Yes
Yes
A009
Transmitter Initializing/Warming Up
Ignore
Yes
No
A010
Calibration Failure
Fault
No
No
A011
Zero Too Low
Fault
Yes
No
A012
Zero Too High
Fault
Yes
No
A013
Zero Too Noisy
Fault
Yes
No
A014
Transmitter Failed
Fault
No
No
A016
Line RTD Temperature Out-of-Range
Fault
Yes
Yes
A017
Meter RTD Temperature Out-of-Range
Fault
Yes
Yes
A020
Calibration Factors Unentered (FlowCal)
Fault
Yes
No
A021
Incorrect Sensor Type (K1)
Fault
No
No
A029
PIC/Daughterboard Communication Failure
Fault
No
No
A030
Incorrect Board Type
Fault
No
No
A031
Low Power
Fault
No
No
A032
Meter Verification/Outputs In Fault
Fault
No
No
A033
Sensor OK, Tubes Stopped by Process
Fault
Yes
Yes
A102
Drive Overrange/Partially Full Tube
Info
Yes
No
A104
Calibration in Progress
Info
Yes
(1)
No
A105
Slug Flow
Info
Yes
No
A107
Power Reset Occurred
Info
Yes
No