Emerson Process Management MICRO MOTION 2400S User Manual
Page 160

152
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Device Profile
15
Alarm status 4
UINT
Get
V
A collection of status bits
• 0x0001 = API:
Temperature out of
range
• 0x0002 = API:
Density out of range
• 0x0004 = Line RTD
out of range
• 0x0008 = Meter RTD
out of range
• 0x0010= Reverse
flow
• 0x0020 = Factory
data error
• 0x0040 = ED: bad
curve
• 0x0080 = LMV
override
• 0x0100 = ED:
Extrapolation error
• 0x0200 = Need
calibration factor
• 0x0400 = Non-volatile
memory error (2700)
• 0x0800 = RAM error
(2700)
• 0x1000= Transmitter
not characterized
• 0x2000 = Non-volatile
memory error (CP)
• 0x4000 = Non-volatile
memory error (CP)
• 0x8000 = Non-volatile
memory error (CP)
16
Alarm status 5
UINT
Get
V
A collection of status bits
• 0x0001 = Boot sector
(CP)
• 0x0002 = Not used
• 0x0004 = Not used
• 0x0008 = Not used
• 0x0010 = Not used
• 0x0020 = Not used
• 0x0040 = Not used
• 0x0080 = Not used
• 0x0100 = D3
calibration in progress
• 0x0200 = D4
calibration in progress
• 0x0400 =
Temperature slope
calibration in progress
• 0x0800 =
Temperature offset
calibration in progress
• 0x1000 = FD
calibration in progress
• 0x2000 = D2
calibration in progress
• 0x4000 = D1
calibration in progress
• 0x8000 = Zero
calibration in progress
Table C-7
Diagnostics Object (0x66) – Instance 1 continued
Attrib
ID
Name
Data type
Service
Mem
Description
Comments