C.8 api object (0x69), Api object (0x69), L, see table c-10 – Emerson Process Management MICRO MOTION 2400S User Manual
Page 170: See table c-10
![background image](/manuals/82275/170/background.png)
162
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Device Profile
C.8
API Object (0x69)
26
Enable offline
password
BOOL
Set
NV
Enable or disable the
password requirement to
access the offline menu
• 0 = Disabled
• 1 = Enabled
27
Offline password
UINT
Set
NV
The offline password for
entering the offline menu
0–9999
28
Update period
UINT
Set
NV
The period in which the
display is updated
Unit = milliseconds
29
Process variable
index
USINT
Set
V
The process variable in
which the precision will be
set in Attribute 30
See Table C-15 for
codes.
30
Process variable
precision
USINT
Set
NV
The number of digits
displayed to the right of
the decimal point for the
process variable selected
with Attribute 29
0–5
31
Language
USINT
Set
NV
Display language
selection
• 0 = English
• 1 = German
• 2 = French
• 3 = Katakana
(1)
• 4 = Spanish
32
Enable IrDA port
USINT
Set
NV
Enable or disable the
IrDA port
• 0 = Disabled
• 1 = Enabled
(1) Not available in this release.
Table C-10
API Object (0x69) – Instance 1
Attrib
ID
Name
Data type
Service
Mem
Description
Comments
1
Temperature
corrected
density
REAL
Get
V
Current value
2
Temperature
corrected
(standard)
volume flow
REAL
Get
V
Current value
3
Temperature
corrected
(standard)
volume total
REAL
Get
Reset
(1)
V
Current value
4
Temperature
corrected
(standard)
volume
inventory
REAL
Get
Reset
(2)
V
Current value
5
Batch weighted
average density
REAL
Get
V
Current value
6
Batch weighted
average
temperature
REAL
Get
V
Current value
Table C-9
Local Display Object (0x68) – Instance 1 continued
Attrib
ID
Name
Data type
Service
Mem
Description
Comments