1 obtaining the test point values, 2 evaluating the test points, Obtaining the test point values – Emerson Process Management MICRO MOTION 2400S User Manual
Page 131: Evaluating the test points

Configuration and Use Manual
123
Troubleshooting
Me
asu
reme
nt P
er
form
anc
e
Defaults
T
rou
b
lesho
o
ti
ng
Compensation
11.19.1
Obtaining the test point values
To obtain the test point values:
•
With the display, configure the required test points as display variables. See Section 8.9.5.
•
With ProLink II
a.
Click
ProLink
>
Diagnostic Information
.
b.
Observe or record the values displayed for
Tube Frequency
,
Left Pickoff
,
Right Pickoff
,
and
Drive Gain
.
•
With a DeviceNet tool, execute a Get for the attributes listed in Table 11-4.
11.19.2
Evaluating the test points
Use the following guidelines to evaluate the test points:
•
If the drive gain is erratic, negative, or saturated, refer to Section 11.19.3.
•
If the value for the left or right pickoff does not equal the appropriate value from Table 11-5,
based on the sensor flow tube frequency, refer to Section 11.19.4.
•
If the values for the left and right pickoffs equal the appropriate values from Table 11-5, based
on the sensor flow tube frequency, record your troubleshooting data and contact Micro Motion
customer service. See Section 11.3.
Table 11-4
Test points with DeviceNet tool
Test point
Class
Instance
Attribute
Drive gain
Diagnostics Object (0x66)
1
20
Tube period
21
Left pickoff
23
Right pickoff
24
Table 11-5
Sensor pickoff values
Sensor
(1)
(1) If your sensor is not listed, contact Micro Motion. See
.
Pickoff value
ELITE
®
CMF sensors
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
F025, F050, F100 sensors
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
F200 sensors
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
H025, H050, H100 sensors
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
H200 sensors
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
R025, R050, or R100 sensors
3.4 mV peak-to-peak per Hz based on sensor flow tube frequency
R200 sensors
2.0 mV peak-to-peak per Hz based on sensor flow tube frequency
T-Series sensors
0.5 mV peak-to-peak per Hz based on sensor flow tube frequency
CMF400 I.S. sensors
2.7 mV peak-to-peak per Hz based on sensor flow tube frequency