3 micro motion customer service, 4 transmitter does not operate, 5 transmitter does not communicate – Emerson Process Management MICRO MOTION 2400S User Manual
Page 120: Micro motion customer service, Transmitter does not operate, Transmitter does not communicate, Section 11.4, Section 11.5
![background image](/manuals/82275/120/background.png)
112
Micro Motion
®
Model 2400S Transmitters for DeviceNet
™
Troubleshooting
11.3
Micro Motion customer service
To speak to a customer service representative, contact the Micro Motion customer service department.
Contact information is provided in Section 1.10.
Before contacting Micro Motion customer service, review the troubleshooting information and
procedures in this chapter, and have the results available for discussion with the technician.
11.4
Transmitter does not operate
If the transmitter is not receiving power, all three LEDs on the user interface will be off.
1. Check the DeviceNet connector (see Section 11.7.1).
2. Ensure that the network is providing sufficient power to the device.
If at least one LED is lit, perform all of the procedures in Section 11.7.
If the procedures do not indicate a problem with the electrical connections, contact the Micro Motion
customer service department.
11.5
Transmitter does not communicate
If the transmitter does not appear to be communicating, the wiring may be faulty or the
communications device may be incompatible. Check the wiring and the communication device. See
Chapter 4 for ProLink II and Pocket ProLink, or Chapter 5 for a DeviceNet tool.
If you are trying to communicate via the IrDA port, ensure that the port is enabled, that read-write
access is enabled, and that there is no active connection via the service port clips. See Section 8.10.6.
Simulation mode for process variables
Checking the flow measurement configuration