Model 2400S Transmitters for DeviceNet
™
Device Profile
55
Meter
verification abort
code
USINT
Get
V
The reason the meter
verification routine
aborted
• 0 = No error
• 1 = Manual abort
• 2 = Watchdog timeout
• 3 = Frequency drift
• 4 = High peak drive
voltage
• 5 = High drive current
standard deviation
• 6 = High drive current
mean
• 7 = Drive loop
reported error
• 8 = High Delta T
standard deviation
• 9 = High Delta T
value
• 10 = State running
• 11 = Verification
complete
• 12 = Wrong
verification enable
• 13 = No factory air
verification
• 14 = No factory water
verification
• 15 = Parameters not
set
56
Meter
verification
algorithm state
at abort
USINT
Get
V
The state of the meter
verification routine when it
aborted
1–18
57
Meter
verification
percent
complete
USINT
Get
V
The progress of the meter
verification routine
%
58
Meter
verification
outputs state
USINT
Set
NV
The state of the outputs
when the meter
verification routine is
running
• 0 = Last value
• 1 = Fault
59
Meter
verification
stiffness limit
REAL
Set
NV
The setpoint of the
stiffness limit. Represents
percentage.
Unitless
60
Meter
verification
validation
counter
UINT
Get
NV
Indicates the number of
times the meter
verification routine has
successfully completed
61
Meter
verification inlet
stiffness out of
limits
USINT
Get
V
Is the inlet stiffness out of
limits?
• 0 = No
• 1 = Yes
62
Meter
verification
outlet stiffness
out of limits
USINT
Get
V
Is the outlet stiffness out
of limits?
• 0 = No
• 1 = Yes
Table C-7
Diagnostics Object (0x66) – Instance 1 continued
Attrib
ID
Name
Data type
Service
Mem
Description
Comments