Checking and adjusting
6
10
12
21
After the first measurement of tilt, the
menu for non-motorized instruments
is the following:
Turn the instrument through 180°
(200 gon) so that
∆
Hz = 0° 00' 00"
(0.0000 gon).
Compensator , continued
If the differences between the hori-
zontal and vertical angles lie within
±4° 30' (±5 gon), the display can be
quitted with
.
The user is made aware by an
acoustic signal that the
-key is
redefined as "OK".
Activates the second tilt
measurement.
Terminates the determination
of the compensator indexes.
The following dialog shows the two
newly-determined values for the
longitudinal and transverse
compensator-index errors.
Stores the new values.
Repeats the complete
calibration procedure.
Leaves the previous values
unchanged.
If the values for the index
errors (l, t) exceed 5' 24"
(0.1 gon), the complete calibration
procedure repeat, but first check that
the instrument is correctly levelled
and is free of vibration. If these
values are exceeded repeatedly,
please notify service.