beautypg.com

Event status enable, Error status register query – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 98

background image

E

VENT

S

TATUS

E

NABLE

Description:

Sets or returns the event status enable bits.

Syntax:

• *ESE

xxx

–or–

• *ESE?

Parameters:

Event status enable bits (xxx):

• Range = 0–255

Refers to the binary word that corresponds to the

event status enable bits.

Example:

• To set the event status enable bits to 10011111

(159 base 10): *ESE 159

• To check the status of the error register enable bits:

*ESE?

The FM5004 will return: *ESE 159

E

RROR

S

TATUS

R

EGISTER

Q

UERY

Description:

Returns the error status register bits.

Syntax:

*ESR?

Parameters:

None

Example:

To check the status of the error status register bits:

*ESR?

The FM5004 will return a number in the range of

0 to 255 that corresponds to the binary representation

of the error status register bits, such as:

*ESR 255

98 |

Remote Operation