Channel select switches – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 27
Switch
Description
Disp
When this LED is illuminated for a selected channel, that channel
is display selected, meaning that the readings from this probe will
be routed to the display. The FM5004 displays the channel input
as a numeric value. To set display mode parameters, use
Menu Select. See Menu Select on page 32 for more information.
Out
When this LED is illuminated for a selected channel, that channel
is analog output selected; meaning that the readings from this
probe will be routed to the analog output system. See ANALOG
OUT SETUP on page 42.
X, Y, Z
This activates a particular axis on selected channel(s) for display
or analog output. Any combination of axes may be chosen;
isotropic response requires selecting all three axes. When multiple
axes are selected, individual axes are combined vectorially.
Some probes do not allow an individual axis to be disabled.
C
HANNEL
S
ELECT
S
WITCHES
The Channel Select Switches
include Ch-1, Ch-2, Ch-3,
and Ch-4.
These switches are used to
select the input channel.
These switches determine the channel for which the base mode settings apply. A
different base mode may be set up for each channel.
Front Panel Controls and Indicators
| 27
Archived 10/19/10