beautypg.com

Channel select switches – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 27

background image

Switch

Description

Disp

When this LED is illuminated for a selected channel, that channel

is display selected, meaning that the readings from this probe will

be routed to the display. The FM5004 displays the channel input

as a numeric value. To set display mode parameters, use

Menu Select. See Menu Select on page 32 for more information.

Out

When this LED is illuminated for a selected channel, that channel

is analog output selected; meaning that the readings from this

probe will be routed to the analog output system. See ANALOG

OUT SETUP on page 42.

X, Y, Z

This activates a particular axis on selected channel(s) for display

or analog output. Any combination of axes may be chosen;

isotropic response requires selecting all three axes. When multiple

axes are selected, individual axes are combined vectorially.

Some probes do not allow an individual axis to be disabled.

C

HANNEL

S

ELECT

S

WITCHES

The Channel Select Switches

include Ch-1, Ch-2, Ch-3,

and Ch-4.

These switches are used to

select the input channel.

These switches determine the channel for which the base mode settings apply. A

different base mode may be set up for each channel.

Front Panel Controls and Indicators

| 27

Archived 10/19/10