beautypg.com

Temp alarm setup – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual

Page 50

background image

Lower Enable:

The alarm will activate when the measured field falls

below the lower alarm value.

Lower Value:

Calls the numeric entry screen to enter the lower

measured field value below which the alarm will

activate.

Latched:

The alarm will sound continuously when an enabled

upper or lower value is crossed, even if the measured

field returns to within the normal field level.

To temporarily disable the alarm, select Alarm Silence

from the Main Menu or power the FM5004 on and off.

TEMP

ALARM

SETUP

Contains the items

associated with

setting up the alarm

for an out of range

temperature

condition.

The temperature alarm will become inactive when the sample rate is set above

10 samples per second. However, taking a temperature measurement from

either the front panel or by a remote temperature measurement command will

cause an alarm if the alarm is enabled and the measurement causes an alarm

condition.

50 |

Menu System

Archived 10/19/10