Error code f: batt fail at fail level, Error code g: double flt–receive–probe reset, Error code h: probe fail–probe did not recover – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 107: Error code l: temp lower–temp under lower limits, Error code n: non support–probe not supported, Error code o: over range error, Error code p: process err no time to process data

E
RROR
C
ODE
F:
B
ATT
F
AIL
A
T
F
AIL
L
EVEL
A probe battery fell below the level at which proper probe operation is
not guaranteed. Recharge the probe battery.
E
RROR
C
ODE
G:
D
OUBLE
F
LT
–R
ECEIVE
–P
ROBE
R
ESET
A probe did not respond for two consecutive communication cycles.
The FM5004 will attempt to reset the probe and re-establish
communication.
E
RROR
C
ODE
H:
P
ROBE
F
AIL
–P
ROBE
D
ID
N
OT
R
ECOVER
A probe ceased to respond to and cannot be reset by the FM5004.
Verify the probe is turned on and the battery is charged. Also, verify
that the fiber optic cables are connected properly.
E
RROR
C
ODE
L:
T
EMP
L
OWER
–T
EMP
U
NDER
L
OWER
L
IMITS
A probe temperature measurement fell below the lower limit set for the
probe temperature alarm.
E
RROR
C
ODE
N:
N
ON
S
UPPORT
–P
ROBE
N
OT
S
UPPORTED
A probe connected to the FM5004 measures field strength in improper
units.
E
RROR
C
ODE
O:
O
VER
R
ANGE
E
RROR
A field strength measurement was above the upper limit for the current
range on the probe being used. When in autorange mode, this error
will occur if the measurement is above the allowable field strength for
the highest range of the probe.
E
RROR
C
ODE
P:
P
ROCESS
E
RR
N
O
T
IME TO
P
ROCESS
D
ATA
The FM5004 did not have time to process the data from a field strength
measurement before another measurement must be taken.
Troubleshooting and Error Handling
| 107