A probe battery fell below the level at which proper probe operation is

not guaranteed. Recharge the probe battery.

E

RROR

C

ODE

G:

D

OUBLE

F

LT

–R

ECEIVE

–P

ROBE

R

ESET

A probe did not respond for two consecutive communication cycles.

The FM5004 will attempt to reset the probe and re-establish

communication.

E

RROR

C

ODE

H:

P

ROBE

F

AIL

–P

ROBE

D

ID

N

OT

R

ECOVER

A probe ceased to respond to and cannot be reset by the FM5004.

Verify the probe is turned on and the battery is charged. Also, verify

that the fiber optic cables are connected properly.

E

RROR

C

ODE

L:

T

EMP

L

OWER

–T

EMP

U

NDER

L

OWER

L

IMITS

A probe temperature measurement fell below the lower limit set for the

probe temperature alarm.

E

RROR

C

ODE

N:

N

ON

S

UPPORT

–P

ROBE

N

OT

S

UPPORTED

A probe connected to the FM5004 measures field strength in improper

units.

E

RROR

C

ODE

O:

O

VER

R

ANGE

E

RROR

A field strength measurement was above the upper limit for the current

range on the probe being used. When in autorange mode, this error

will occur if the measurement is above the allowable field strength for

the highest range of the probe.

E

RROR

C

ODE

P:

P

ROCESS

E

RR

N

O

T

IME TO

P

ROCESS

D

ATA

The FM5004 did not have time to process the data from a field strength

measurement before another measurement must be taken.