Display update rate, Lcd contrast, Lcd brightness – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 41

DISPLAY
UPDATE
RATE
Sets the display
update rates, which
is the number of
times per second
that the display will
update with new
readings.
If the samples per second setting in the Main Menu is greater than the display
update rate, then all samples taken after the last display update will be included
in the new update. The displayed reading will be an average of the previous
samples.
In 2-line display mode, with minimum or maximum of probe selected, the
displayed reading will be the minimum or maximum of the previous samples. If
the display update rate is set greater than samples per second, the display will
update only when new samples are taken.
LCD
CONTRAST
Adjusts the display contrast. This can make the display more readable under
different lighting conditions and viewing angles.
Improper adjustment of the contrast may make the display
unreadable. If the display is blank, the contrast may be set too low. If
the display is completely dark, the contrast may be set too high.
LCD
BRIGHTNESS
Adjusts the brightness of the display. This can make the display more readable
under different lighting conditions and viewing angles.
Menu System
| 41
Archived 10/19/10